Recall Z-2507-2026
- Recall number
- Z-2507-2026
- Event number
- 98994
- Firm FEI
- 1417592
- Product code
- OJK
- Status
- Open, Classified
- Initiated
- 2026-05-04
- Posted
- 2026-06-17
- Root cause
- Nonconforming Material/Component
Product#
MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775K; 4) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775L; 5) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775M; 6) TONSIL PACK, Medline Kit Number/SKU DYNJ84480; 7) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080; 8) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080A.
Reason for Recall#
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.