ASTM F2554-10

Standard Organization:

ASTM F2554-10
Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems

Recognition Date2013-01-15
Recognition List030
Recognition Number11-251
Extent Complete standard
Standards Orgs ASTM ASTM International http://www.astm.org/
STG Orthopedic
FDA Tech .stds { color: #000000; float: center; width: 950px; margin: 0px 25px 0px 0px; } .stds-middle { color: #000000; float: center; width: 575px; margin: 0px 50px 50px 50px; } /* .stds-left { float: left; margin: 9px 30px 0px 0; } */ .stds-left { color: #000000; float: left; width: 700px; margin: 9px 0px 0px 0px; } .stds-search-left { float: left; width: 610px; margin: 9px 30px 0px 0; } .stds-search-right { float: right; width: 190px; margin: 9px 18px 0 0; padding: 8px 12px 10px 12px; background-color: #f9f9f3; border: 1px solid #b0b0b0; -webkit-border-radius: 6px; -moz-border-radius: 6px; border-radius: 6px; } .stds-right { color: #000000; float: right; width: 170px; margin: 9px 18px 0 0; padding: 8px 12px 10px 12px; background-color: #f3f3f3; border: 1px solid #b0b0b0; border-radius: 6px; } .stds-intro { margin: 0px 0px 18px 0px; padding: 5px 12px 12px 12px; background-color: #e8eff5; border: 1px solid #b0b0b0; border-radius: 6px; } #stds_nr-form { margin-top: 18px; padding-top: 12px; border-top: 1px dashed #b0b0b0; } #stds-form { table-layout:fixed; width: 700px; font-size: 11px; border: 1px solid #b0b0b0; background-color: #f9f9f3; } #stds-form-title{ padding: 3px 3px 3px 3px; width: 100%; background-color: #e8eff5; border-bottom: 1px solid #b0b0b0; margin-top: 0px; } #stds-form-title-left { font-size: 13px; font-weight:normal; } #stds-form-title-right { /*float: right;*/ font-size: 11px; padding: 0px 0px 0px 500px; } #stds-form-results-right { /*float: right;*/ font-size: 11px; padding: 0px 0px 0px 600px; } #stds-search { height: 18px; margin: 0px; padding: 3px 10px; border: 1px solid #b0b0b0; font-size: 12px; color: #484138; } #stds-search:focus, #stds-search:hover { -webkit-box-shadow: 0 0px 6px rgba(0,0,0,.1); -moz-box-shadow: 0 0px 6px rgba(0,0,0,.1); box-shadow: 0 0px 6px rgba(0,0,0,.1); } #stds-results-number { margin-bottom: 12px; font-size: 10px; font-weight: bold; } #stds-results-table { margin-left:auto; margin-right:auto; text-align:left; padding: 0; border-collapse:collapse; table-layout:fixed; background-color: #f9f9f3; border-color: #b0b0b0; border-width: 0 0 1px 1px; border-style: solid; } #stds-results-table th { padding: 6px 12px; background-color: #f9f9f3; border-color: #b0b0b0; border-width: 1px 1px 0 0; border-style: solid; font-size: 12px; } #stds-results-table td { padding: 6px 12px; border-color: #b0b0b0; border-width: 1px 1px 0 0; border-style: solid; text-transform: capitalize; } #stds-results-table tr:nth-child(odd) { /* background-color: #f3f3f3; */ } .sort { float: right; margin: 0px 10px 0px 0px; padding: 0px 5px 5px 5px; } /* .sort { float: right; margin-top: 5px;} */ .sort a { color: #484138; } .sort a:hover { color: #0f2a8c; } Recognized Consensus Standards // Quick Links: Skip to main page content Skip to Search Skip to Topics Menu Skip to Common Links U.S. Food & Drug Administration   Follow FDA En Español Search FDA   Home Food Drugs Medical Devices Radiation-Emitting Products Vaccines, Blood & Biologics Animal & Veterinary Cosmetics Tobacco Products   Recognized Consensus Standards var MP = { Version: '1.0.22', Domains: {'es':'esp.fda.gov'}, SrcLang: 'en', UrlLang: 'mp_js_current_lang', SrcUrl: unescape('mp_js_orgin_url'), init: function(){ if (MP.UrlLang.indexOf('p_js_')==1) { MP.SrcUrl=window.top.document.location.href; MP.UrlLang=MP.SrcLang; } }, getCookie: function(name){ var start=document.cookie.indexOf(name+'='); if(start < 0) return null; start=start+name.length+1; var end=document.cookie.indexOf(';', start); if(end < 0) end=document.cookie.length; while (document.cookie.charAt(start)==' '){ start++; } return unescape(document.cookie.substring(start,end)); }, setCookie: function(name,value,path,domain){ var cookie=name+'='+escape(value); if(path)cookie+='; path='+path; if(domain)cookie+='; domain='+domain; var now=new Date(); now.setTime(now.getTime()+1000*60*60*24*365); cookie+='; expires='+now.toUTCString(); document.cookie=cookie; }, switchLanguage: function(lang){ if(lang!=MP.SrcLang){ var script=document.createElement('SCRIPT'); script.src=location.protocol+'//'+MP.Domains[lang]+'/'+MP.SrcLang+lang+'/?1023749632;'+encodeURIComponent(MP.SrcUrl); document.body.appendChild(script); } else if(lang==MP.SrcLang && MP.UrlLang!=MP.SrcLang){ var script=document.createElement('SCRIPT'); script.src=location.protocol+'//'+MP.Domains[MP.UrlLang]+'/'+MP.SrcLang+MP.UrlLang+'/?1023749634;'+encodeURIComponent(location.href); document.body.appendChild(script); } return false; }, switchToLang: function(url) { window.top.location.href=url; } }; var addthis_config = {data_track_clickback: true, ui_508_compliant: true, services_compact: 'facebook, twitter, linkedin, pinterest_share, email', services_exclude: 'print', image_container: "#user_provided" } ; FDA Home Medical Devices Databases -

Rationale:

This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.

Scope:

1.1 This practice addresses the techniques of measurement and reporting of basic static performance (accuracy, repeatability, and so forth) of surgical navigation and/or robotic positioning devices under defined conditions. The scope covers the tracking subsystem, testing only in this practice the accuracy and repeatability of the system to locate individual points in space. A point in space has no orientation; only multi-dimensional objects have orientation. Therefore, orientation of objects is not within the scope of this practice. However, in localizing a point the different orientations of the localization tool can produce errors. These errors and the orientation of the localization tool are within the scope of this practice. The aim is to provide a standardized measurement of performance variables by which end-users can compare within (for example, different fixed reference frames or stylus tools) and between (for example, different manufacturers) different systems. Parameters to be evaluated include (based upon the features of the system being evaluated):

(1) Location of a point relative to a coordinate system.

(2) Relative point to point accuracy (linear).

(3) Repeatability of coordinates of a single point.

(4) For an optically based system, the range of visible orientations of the reference frames or tools.

(5) This method covers all configurations of tool arrays in the system.

1.2 The system as defined in this practice includes only the tracking subsystem (optical, magnetic, mechanical, and so forth) stylus, computer, and necessary hardware and software. As such, this practice incorporates tests that can be applied to a prescribed phantom model in a laboratory or controlled setting.

1.3 This practice defines a standardized reporting format, which includes definition of the coordinate systems to be used for reporting the measurements, and statistical measures (for example, mean, standard deviation, maximum error).

1.4 This practice will serve as the basis for subsequent standards for specific tasks (cutting, drilling, milling, reaming, biopsy needle placement, and so forth) and surgical applications.

Code of Federal Regulations:

Regulation Number Device Name Device Class Product Code
§882.4560 Neurological Stereotaxic Instrument, Real-Time Intraoperative Mri Class 2 ORR
§882.4560 Orthopedic Stereotaxic Instrument Class 2 OLO
§882.4560 Patient Specific Manual Orthopedic Stereotaxic System Class 2 OSF

FDA Guidance:

There is no relevant guidance developed at this time.