CLSI H26-A2 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Proposed Standard - Second Edition |
Recognition Date | 2014-01-30 |
Recognition List | 034 |
Recognition Number | 7-210 |
Extent | Partial recognition. The following part(s) of the standard is (are) not recognized Section 5.5.7 Medically Allowable Error Section 5.6.1 Instrument Calibration Section 5.6.2Whole Blood Cross-Checks |
Standards Orgs | CLSI Clinical Laboratory Standards Institute https://clsi.org/ |
STG | InVitro Diagnostics |
FDA Tech | Takeesha Taylor-Bell FDA/OMPT/CDRH/OIR/DIHD/HEMB/ 240-402-6566 Takeesha.Taylor-Bell@fda.hhs.gov |
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. Section 5.5.7 is in conflict with methods to estimate bias and confidence intervals, see article listed below. Section 5.6.1 is in conflict with 21 CFR 820.72 (b)(1), see the regulation listed below. Section 5.6.2 is in conflict with 21 CFR 820.72 (b)(1), see the regulation listed below.
This document provides guidance for the validation, verification, calibration, quality assurance (QA), and quality control (QC) of automated multichannel hematology analyzers for manufacturers, end-user clinical laboratories, accrediting organizations, and regulatory bodies. In addition, end-user clinical laboratories will find guidance for establishment of clinically reportable intervals and for QA for preexamination and examination aspects of their systems.
Regulation Number | Device Name | Device Class | Product Code |
§864.5200 | Counter, Cell, Automated (Particle Counter) | Class 2 | GKL |
§864.5200 | Instrument, Automated Platelet Counting | Class 2 | GKX |
§864.5200 | Counter, Urine Particle | Class 2 | LKM |
§864.5200 | Counter, Cell, Automated (Particle Counter) | Class 2 | GKL |
§864.5200 | Counter, Urine Particle | Class 2 | LKM |
§864.5200 | Instrument, Automated Platelet Counting | Class 2 | GKX |
§864.5220 | Counter, Differential Cell | Class 2 | GKZ |
Martin, R.L. (2000). General Deming Regression for Estimating Systematic Bias and Its Confidence Intervals in Method-Comparison Studies. Clinical Chemistry, 46(1):100-104. |