ASTM F3141-15

Standard Organization:

ASTM F3141-15
Standard Guide for Total Knee Replacement Loading Profiles

Recognition Date2016-12-23
Recognition List046
Recognition Number11-318
Extent Complete standard
Standards Orgs ASTM ASTM International http://www.astm.org/
STG Orthopedic
FDA Tech .stds { color: #000000; float: center; width: 950px; margin: 0px 25px 0px 0px; } .stds-middle { color: #000000; float: center; width: 575px; margin: 0px 50px 50px 50px; } /* .stds-left { float: left; margin: 9px 30px 0px 0; } */ .stds-left { color: #000000; float: left; width: 700px; margin: 9px 0px 0px 0px; } .stds-search-left { float: left; width: 610px; margin: 9px 30px 0px 0; } .stds-search-right { float: right; width: 190px; margin: 9px 18px 0 0; padding: 8px 12px 10px 12px; background-color: #f9f9f3; border: 1px solid #b0b0b0; -webkit-border-radius: 6px; -moz-border-radius: 6px; border-radius: 6px; } .stds-right { color: #000000; float: right; width: 170px; margin: 9px 18px 0 0; padding: 8px 12px 10px 12px; background-color: #f3f3f3; border: 1px solid #b0b0b0; border-radius: 6px; } .stds-intro { margin: 0px 0px 18px 0px; padding: 5px 12px 12px 12px; background-color: #e8eff5; border: 1px solid #b0b0b0; border-radius: 6px; } #stds_nr-form { margin-top: 18px; padding-top: 12px; border-top: 1px dashed #b0b0b0; } #stds-form { table-layout:fixed; width: 700px; font-size: 11px; border: 1px solid #b0b0b0; background-color: #f9f9f3; } #stds-form-title{ padding: 3px 3px 3px 3px; width: 100%; background-color: #e8eff5; border-bottom: 1px solid #b0b0b0; margin-top: 0px; } #stds-form-title-left { font-size: 13px; font-weight:normal; } #stds-form-title-right { /*float: right;*/ font-size: 11px; padding: 0px 0px 0px 500px; } #stds-form-results-right { /*float: right;*/ font-size: 11px; padding: 0px 0px 0px 600px; } #stds-search { height: 18px; margin: 0px; padding: 3px 10px; border: 1px solid #b0b0b0; font-size: 12px; color: #484138; } #stds-search:focus, #stds-search:hover { -webkit-box-shadow: 0 0px 6px rgba(0,0,0,.1); -moz-box-shadow: 0 0px 6px rgba(0,0,0,.1); box-shadow: 0 0px 6px rgba(0,0,0,.1); } #stds-results-number { margin-bottom: 12px; font-size: 10px; font-weight: bold; } #stds-results-table { margin-left:auto; margin-right:auto; text-align:left; padding: 0; border-collapse:collapse; table-layout:fixed; background-color: #f9f9f3; border-color: #b0b0b0; border-width: 0 0 1px 1px; border-style: solid; } #stds-results-table th { padding: 6px 12px; background-color: #f9f9f3; border-color: #b0b0b0; border-width: 1px 1px 0 0; border-style: solid; font-size: 12px; } #stds-results-table td { padding: 6px 12px; border-color: #b0b0b0; border-width: 1px 1px 0 0; border-style: solid; text-transform: capitalize; } #stds-results-table tr:nth-child(odd) { /* background-color: #f3f3f3; */ } .sort { float: right; margin: 0px 10px 0px 0px; padding: 0px 5px 5px 5px; } /* .sort { float: right; margin-top: 5px;} */ .sort a { color: #484138; } .sort a:hover { color: #0f2a8c; } Recognized Consensus Standards // Quick Links: Skip to main page content Skip to Search Skip to Topics Menu Skip to Common Links U.S. Food & Drug Administration   Follow FDA En Español Search FDA   Home Food Drugs Medical Devices Radiation-Emitting Products Vaccines, Blood & Biologics Animal & Veterinary Cosmetics Tobacco Products   Recognized Consensus Standards var MP = { Version: '1.0.22', Domains: {'es':'esp.fda.gov'}, SrcLang: 'en', UrlLang: 'mp_js_current_lang', SrcUrl: unescape('mp_js_orgin_url'), init: function(){ if (MP.UrlLang.indexOf('p_js_')==1) { MP.SrcUrl=window.top.document.location.href; MP.UrlLang=MP.SrcLang; } }, getCookie: function(name){ var start=document.cookie.indexOf(name+'='); if(start < 0) return null; start=start+name.length+1; var end=document.cookie.indexOf(';', start); if(end < 0) end=document.cookie.length; while (document.cookie.charAt(start)==' '){ start++; } return unescape(document.cookie.substring(start,end)); }, setCookie: function(name,value,path,domain){ var cookie=name+'='+escape(value); if(path)cookie+='; path='+path; if(domain)cookie+='; domain='+domain; var now=new Date(); now.setTime(now.getTime()+1000*60*60*24*365); cookie+='; expires='+now.toUTCString(); document.cookie=cookie; }, switchLanguage: function(lang){ if(lang!=MP.SrcLang){ var script=document.createElement('SCRIPT'); script.src=location.protocol+'//'+MP.Domains[lang]+'/'+MP.SrcLang+lang+'/?1023749632;'+encodeURIComponent(MP.SrcUrl); document.body.appendChild(script); } else if(lang==MP.SrcLang && MP.UrlLang!=MP.SrcLang){ var script=document.createElement('SCRIPT'); script.src=location.protocol+'//'+MP.Domains[MP.UrlLang]+'/'+MP.SrcLang+MP.UrlLang+'/?1023749634;'+encodeURIComponent(location.href); document.body.appendChild(script); } return false; }, switchToLang: function(url) { window.top.location.href=url; } }; var addthis_config = {data_track_clickback: true, ui_508_compliant: true, services_compact: 'facebook, twitter, linkedin, pinterest_share, email', services_exclude: 'print', image_container: "#user_provided" } ; FDA Home Medical Devices Databases -

Rationale:

This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.

Scope:

1.1 Motion path, load history and loading modalities all contribute to the wear, degradation and damage of implanted prosthetics. Simulating a variety of functional activities promises more realistic testing for wear and damage mode evaluation. Such activities are often called activities of daily living (ADLs). ADLs identified in the literature include walking, stair ascent and decent, sit-to-stand, stand-to-sit, squatting, kneeling, cross-legged sitting, into bath, out of bath, turning and cutting motions (1-7).2 Activities other than walking gait often involve an extended range of motion and higher imposed loading conditions which have the ability to cause damage and modes of failure other than normal wear (8-10).

1.2 This document provides guidance for functional simulation to evaluate the durability of knee prosthetic devices under force control.

1.3 Function simulation is defined as the reproduction of loads and motions that might be encountered in activities of daily living but it does not necessarily cover every possible type of loading. Functional simulation differs from typical wear testing in that it attempts to exercise the prosthetic device through a variety of loading and motion conditions such as might be encountered in situ in the human body in order to reveal various damage modes and damage mechanisms that might be encountered throughout the life of the prosthetic device.

1.4 Force control is defined as the mode of control of the test machine that accepts a force level as the set point input and which utilizes a force feedback signal in a control loop to achieve that set point input. For knee simulation, the flexion motion is placed under angular displacement control, internal and external rotation is placed under torque control, and axial load, anterior posterior shear and medial lateral shear are placed under force control.

1.5 This document establishes kinetic and kinematic test conditions for several activities of daily living, including walking, turning navigational movements, stair climbing, stair descent, and squatting. The kinetic and kinematic test conditions are expressed as reference waveforms used to drive the relevant simulator machine actuators. These waveforms represent motion, as in the case of flexion extension, or kinetic signals representing the forces and moments resulting from body dynamics, gravitation and the active musculature acting across the knee.

1.6 This document does not address the assessment or measurement of damage modes, or wear or failure of the prosthetic device.

1.7 This document is a guide. As defined by ASTM in their "Form and Style for ASTM Standards" book in section C15.2, "A standard guide is a compendium of information or series of options that does not recommend a specific course of action. Guides are intended to increase the awareness of information and approaches in a given subject area. Guides may propose a series of options or instructions that offer direction without recommending a definite course of action. The purpose of this type of standard is to offer guidance based on a consensus of viewpoints but not to establish a standard practice to follow in all cases." The intent of this guide is to provide loading profiles and test procedures to develop testing that might be used for wear, durability or other types of testing of total knee replacements. As noted in this definition, a guide provides guidance on testing, but does not require specific testing. Thus, for example, if a user is unable to control one mode of force control given in the load profiles, that user is not required to perform that mode of loading.

Code of Federal Regulations:

Regulation Number Device Name Device Class Product Code
§888.3565 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 2 MBH
§888.3560 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uhmwpe, Pegged, Cemented, Polymer/Metal/Polymer Class 2 MBV
§888.3560 Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive Class 2 OIY
§888.3560 Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer Class 2 JWH
§888.3550 Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Metal Class 3 KRP
§888.3550 Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Polymer Class 2 KRQ
§888.3530 Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer Class 2 HRY
§888.3520 Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer Class 2 HSX
§888.3510 Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer Class 2 KRO
§888.3500 Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Composite Class 2 KYK
§888.3490 Prosthesis, Knee, Femorotibial, Non-Constrained, Metal/Composite Cemented Class 2 KTX
§888.3490 Prosthesis, Knee, Non-Constrained (Metal-Carbon Reinforced Polyethylene) Cemented Class 2 KMB
§888.3480 Metal Cemented Constrained Femorotibial Knee Prosthesis Class 3 KRN
Unclassified Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Polymer/Metal/Polymer Class 3 LXY
Unclassified Prosthesis, Knee, Patello/Femorotibial, Unconstrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 3 MBD
Unclassified Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing Class 3 NJL

FDA Guidance:

Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, January 16, 2003


© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.