Rx only
Contains Y-Sampling Site for the collection of unanticoagulated whole blood samples for laboratory testing and the DonorCare Needle Guard.
Use aseptic technique.
This product has components which contain natural rubber latex.
Note: Nominal tubing dimensions of product are 0.118" inner diameter x 0.025" wall thickness.
Note: If the Y-Sampling Site is not used, donor samples may be collected using an alternate method following standard procedures.
Precautions:
Precaution: Do not proceed with the remaining steps until the entire whole blood unit is collected.
Precaution: Complete steps 13 - 21 within approximately 4 minutes after sealing the donor tubing to avoid possible clot formation in the tubing.
Note: If the access device is assembled such that the outer barrel is screwed onto the Luer, make sure to rotate clockwise upon insertion to avoid barrel detaching from Luer.
Notes:
Precaution: When replacing the access device, be careful to avoid contact with any blood droplets on the Luer or Sampling Site. Discard used access device appropriately.
Precaution: The needle guard must be held stationary while the needle is withdrawn into it.
Note: Step 22 may be performed prior to step 21 if desired.
Baxter Healthcare Corporation
Deerfield, IL 60015 USA
Made in USA
07-19-50-479
Iss. March 2006
Baxter, Blood-Pack and Transfer Pack are trademarks of Baxter International Inc.
Baxter and Blood-Pack are registered in the U.S. Patent and Trademark Office.
DonorCare is a trademark of ITL Corporation
U.S. Patent Nos.: 5,314,421; 5,372,143; 5,507,525
Rx only
This product has components which contain natural rubber latex.
Precaution: Proper handling of segments is critical to assuring component traceability.
The enclosed Blood-Pack unit is manufactured with two different sets of segment numbers as shown below. The use of two different sets of numbers requires special processing in order to assure component traceability through both sets of numbers. The following process is one suggested method which may be implemented in the blood processing lab. This procedure is not intended for use at the whole blood collection site.
Precaution: Be sure to leave the first segment, below Site A, attached to the red cell product along with the donor segments for cross matching.
Precaution: At least one segment must remain attached to all Transfer Pack Containers to ensure traceability.
Notes:
Baxter Healthcare Corporation
Deerfield, IL 60015 USA
Made in USA
07-19-44-336
Iss. August 2004
Baxter, Blood-Pack and Transfer Pack are trademarks of Baxter International Inc.
Code 4R3648
20 Units
Baxter
Anticoagulant Citrate Phosphate Dextrose Adenine Solution, USP (CPDA-1) Blood-Pack Unit
Triple For the Collection and Processing of 450 mL Blood
Y-Sampling Site, DonorCare Needle Guard, 16 ga. Ultra Thin Wall Needle
Each unit consists of a PL 146 Plastic primary container with 63 mL of CPDA-1 solution containing 2 g Dextrose (monohydrate) USP, 1.66 g Sodium Citrate (dihydrate) USP, 188 mg Citric Acid (anhydrous) USP, 140 mg Monobasic Sodium Phosphate (monohydrate) USP and 17.3 mg Adenine USP, pH may have been adjusted with sodium hydroxide; two empty 400 mL PL 1240 Plastic Transfer Pack Containers.
Rx only
Sterile, nonpyrogenic fluid path
See instructions for use.
Store at room temperature.
U.S. Pat. Nos. 5,314,421; 5,372,143; 5,507,525; pat. pend.
Baxter, Blood-Pack, PL 146 and Transfer Pack are trademarks of Baxter International Inc.; Baxter and Blood-Pack are registered in the U.S. Patent and Trademark Office.
DonorCare is a trademark of ITL Corporation
Baxter Healthcare Corporation
Deerfield, IL 60015 USA
Made in USA
07-28-52-752 02/2007
CPDA-1
anticoagulant citrate phosphate dextrose adenine solution |
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
Labeler - Fenwal, Inc. (794519020) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Fenwal International, Inc. | 091164590 | MANUFACTURE(0942-6319) |