Vizamyl by is a Prescription medication manufactured, distributed, or labeled by Medi-Physics, Inc. dba GE Healthcare, Cardinal Health 414, LLC (Sacramento, CA), Cardinal Health 414, LLC (Phoenix, AZ), Precision Nuclear, LLC (Johnson City, TN), Cardinal Health, Inc. (East Lansing, MI), Cardinal Health 414, LLC (Colton, CA), Cardinal Health 414, LLC (Beltsville, MD), Cardinal Health 414, LLC (Seattle, WA), Cardinal Health 414, LLC (Dallas, TX), Cardinal Health 414, LLC (Tampa, FL), Cardinal Health 414, LLC (Charlotte, NC), Cardinal Health 418, Inc. (Aurora, CO), Cardinal Health, Inc. (Houston, TX), Cardinal Health, Inc. (Glendale Heights, IL), Cardinal Health, Inc. (New Orleans, LA), GE Healthcare AS (Oslo, Norway), The University of Utah DBA Cyclotron Radiochemistry Lab Huntsman Cancer Institute (Salt Lake City, UT), Essential Isotopes, LLC. (Colombia, MO). Drug facts, warnings, and ingredients follow.
Vizamyl is a radioactive diagnostic agent indicated for Positron Emission Tomography (PET) imaging of the brain to estimate β-amyloid neuritic plaque density in adult patients with cognitive impairment who are being evaluated for Alzheimer's disease (AD) or other causes of cognitive decline. A negative Vizamyl scan indicates sparse to no neuritic plaques, and is inconsistent with a neuropathological diagnosis of AD at the time of image acquisition; a negative scan result reduces the likelihood that a patient's cognitive impairment is due to AD. A positive Vizamyl scan indicates moderate to frequent amyloid neuritic plaques; neuropathological examination has shown this amount of neuritic plaque is present in patients with AD, but may also be present in patients with other types of neurologic conditions, as well as older people with normal cognition. Vizamyl is an adjunct to other diagnostic evaluations (1).
Limitations of Use:
Injection: 150 MBq/mL (4.05 mCi/mL) at reference date and time in 30 mL multi-dose vials (3).
Known hypersensitivity to Vizamyl or any excipient, including polysorbate 80 (4).
Most commonly reported adverse reactions were flushing (2%), headache (1%), increased blood pressure (2%), nausea (1%), and dizziness (1%).
To report SUSPECTED ADVERSE REACTIONS, contact GE Healthcare at 1-800-654-0118 or FDA at 1-800-FDA-1088 or www.fda.gov/medwatch. (6)
Lactation: A lactating woman may pump and discard breast milk for 24 hours after Vizamyl administration. (8.2)
See 17 for PATIENT COUNSELING INFORMATION.
Revised: 1/2020
Vizamyl is indicated for Positron Emission Tomography (PET) imaging of the brain to estimate β-amyloid neuritic plaque density in adult patients with cognitive impairment who are being evaluated for Alzheimer's Disease (AD) and other causes of cognitive decline.
A negative Vizamyl scan indicates sparse to no neuritic plaques and is inconsistent with a neuropathological diagnosis of AD at the time of image acquisition; a negative scan result reduces the likelihood that a patient's cognitive impairment is due to AD. A positive Vizamyl scan indicates moderate to frequent amyloid neuritic plaques; neuropathological examination has shown this amount of amyloid neuritic plaque is present in patients with AD, but may also be present in patients with other types of neurologic conditions as well as in older people with normal cognition. Vizamyl is an adjunct to other diagnostic evaluations.
Limitations of Use:
Vizamyl is a radioactive drug and should be handled with safety measures to minimize radiation exposure during administration [see Warnings and Precautions (5.3)]. Use waterproof gloves and effective shielding, including lead-glass syringe shields when handling and administering Vizamyl. To minimize radiation dose to the bladder, encourage patients to hydrate before and after Vizamyl administration in order to permit frequent voiding. Encourage patients to void before and after imaging with Vizamyl and frequently thereafter for 24 hours following Vizamyl administration.
Radiopharmaceuticals, including Vizamyl, should be used by or under the control of physicians who are qualified by specific training and experienced in the safe use and handling of radioactive materials, and whose experience and training have been approved by the appropriate governmental agency authorized to license the use of radiopharmaceuticals.
The recommended dose for Vizamyl is 185 megabecquerels (MBq) [5 millicuries (mCi)] in a maximum dose volume of 10 mL, administered as a single intravenous bolus within 40 seconds. The maximum mass dose is 20 micrograms. Follow the injection with an intravenous flush of 5 to 15 mL of 0.9% sterile sodium chloride injection.
The recommended PET scan start time is 60 to 120 minutes after Vizamyl injection, using a PET scanner in 3-D mode with appropriate data corrections. A scan duration of 10 to 20 minutes is recommended. The time of initiation and the duration of the scan may vary depending on dose, imaging acquisition, and reconstruction parameters. Position the patient supine with the brain (including the cerebellum) within a single field of view. The patient's head should be tilted so that the anterior commissure-posterior commissure (AC-PC) plane is at right angles to the bore-axis of the PET scanner, with the head positioned in a suitable head support. Reducing head movement with tape or other flexible head restraints may be employed. Iterative or filtered back-projection reconstruction is recommended with a slice thickness of 2 to 4 mm, matrix size of 128 × 128 with pixel sizes of approximately 2 mm. Where a post-smoothing filter is applied, a full width half maximum (FWHM) of not more than 5 mm is recommended; filter FWHM should be chosen to optimize the signal-to-noise ratio while preserving the sharpness of the reconstructed image.
Image Orientation
Orient axial and coronal images to show symmetry of brain structures, with equal heights of structures bilaterally. Orient sagittal images so that the head and neck are neither flexed nor extended; the anterior and posterior aspects of the corpus callosum should be parallel to the AC-PC line as shown in Figure 2.
Image Display
Vizamyl images should be interpreted only by readers who successfully complete the electronic or in-person training program provided by the manufacturer [see Warnings and Precautions (5.2)]. The objective of Vizamyl image interpretation is to provide an estimate of the brain β-amyloid neuritic plaque density, not to make a clinical diagnosis. Image interpretation is performed independently of a patient's clinical features and relies upon recognition of image features in certain brain regions.
Image interpretation is based upon the distribution of radioactive signal within the brain; clinical information is not a component of image assessment [see Warnings and Precautions (5.2)]. Images are designated as positive or negative either by comparing radioactivity in cortical grey matter with activity in adjacent white matter, or based on the intensity in the five regions mentioned above. Signal uptake in the cerebellum does not contribute to scan interpretation (for example, a positive scan may show retained cerebellar grey-white contrast even when the cortical grey-white contrast is lost). Images should be viewed with the minimum image intensity set to zero and the maximum set such that the signal level in the easily identifiable pons is at 90% of maximum.
Negative scans show more radioactivity in white matter than in grey matter, creating clear grey-white matter contrast.
Specifically, a negative scan would have the following characteristics:
Positive scans show at least one cortical region with reduction or loss of the normally distinct grey-white matter contrast. These scans have one or more regions with increased cortical grey matter signal (above 50-60% peak intensity) and/or reduced (or absent) grey-white matter contrast (white matter sulcal pattern is less distinct). A positive scan may have one or more regions in which grey matter radioactivity is as intense or exceeds the intensity in adjacent white matter.
Specifically, a positive scan would have the following characteristics:
Among patients with clinically important β-amyloid neuritic plaques in the brain, the temporal lobes, parietal lobes, and striatum may not be as affected compared to other brain regions. Therefore, in some images, flutemetamol F 18 signal in these regions may not be as intense as in the frontal lobes or the posterior cingulate and precuneus regions.
Atrophy may affect the interpretability of scans, particularly in the frontal, temporal and parietal lobes [see Warnings and Precautions (5.2)]. For cases in which atrophy is apparent or suspected and there is uncertainty as to the location of the grey matter on the PET scan, examine the striatum for flutemetamol F 18 signal as it is less affected by atrophy than other regions of the brain.
If the patient's MRI or CT brain images are available the interpreter should examine the CT or MRI images to clarify the relationship between PET flutemetamol F 18 uptake and grey matter anatomy.
Other factors that may affect the ability to interpret Vizamyl images include patient factors such as brain pathology, surgical changes, post-radiation therapy changes, and implants. Some scans may be difficult to interpret due to image noise, suboptimal patient positioning, or over-smoothing of the reconstructed image.
Figure 1: Axial view of negative (left) and positive (right) Vizamyl scans. The axial slices which cut through the frontal pole and inferior aspect of the splenium are shown using a rainbow color scale. The left image shows a white matter sulcal pattern at the frontal (f) and lateral temporal (lt) regions with a color intensity that tapers to the periphery, as well as less radioactivity in the striatal region(s). The right image shows absence of the white matter sulcal pattern with intensity radiating to a sharply defined convex edge, as well as more radioactivity in the striatum. In both the frontal and lateral temporal regions, the intensity is higher in the grey matter regions of the right image compared to those of the left image.
Figure 2: Sagittal view of negative (left) and positive (right) Vizamyl scans. The sagittal slices are slightly off midline in one hemisphere and shown using a rainbow color scale. In the posterior cingulate (pc) region, which is superior and posterior to the corpus callosum (cc), the left image shows intensity below 50% of peak intensity whereas the right image shows intensity above 60% of peak intensity. The pons (p) is set to approximately 90% of the maximum intensity.
Figure 3: Coronal view of negative (left) and positive (right) Vizamyl scans. The coronal slices are located posterior to the corpus callosum. The left image shows a white matter sulcal pattern in the inferior parietal (ip) regions that is not evident in the right image. Relative to the left image, the right image shows increased intensity in the posterior cinguli (pc) and increased radial extent of high intensity to the lateral surfaces of the parietal lobes particularly evident in the inferior parietal regions.
The estimated absorbed radiation doses for adult patients following intravenous injection of Vizamyl are shown in Table 1. Values were calculated from human biodistribution data using OLINDA/EXM software and assuming emptying of the urinary bladder at 3.5 hour intervals.
The adult effective dose resulting from a 185-MBq (5-mCi) Vizamyl administration is 5.92 mSv. The use of a CT scan to calculate attenuation correction for reconstruction of Vizamyl images (as done in PET/CT imaging) will add radiation exposure at the level of approximately 0.1 mSv effective dose. Diagnostic head CT scans using helical scanners administer an average of 2.2 ± 1.3 mSv effective dose. The actual radiation dose is operator and scanner dependent.
Organ/Tissue | Absorbed Radiation Dose Per Unit Administered Activity microGy/MBq |
---|---|
Adrenals | 13 |
Brain | 11 |
Breasts | 5 |
Gallbladder wall | 287 |
Heart wall | 14 |
Kidneys | 31 |
Liver | 57 |
Lower large intestine wall | 42 |
Lungs | 16 |
Muscle | 9 |
Osteogenic cells | 11 |
Ovaries | 25 |
Pancreas | 15 |
Red marrow | 13 |
Skin | 5 |
Small intestine wall | 102 |
Spleen | 15 |
Stomach wall | 12 |
Testes | 8 |
Thymus | 6 |
Thyroid | 6 |
Upper large intestine wall | 117 |
Urinary bladder wall | 145 |
Uterus | 25 |
Total body | 12 |
Effective Dose | 32 (microSv/MBq) |
Vizamyl is contraindicated in patients with a history of hypersensitivity reaction to Vizamyl, polysorbate 80, or any other inactive ingredient in Vizamyl [see Warnings and Precautions (5.1)].
Hypersensitivity reactions such as flushing and dyspnea have been observed within minutes following Vizamyl administration. These reactions may occur in patients with no history of prior exposure to Vizamyl.
Before administering Vizamyl, ask patients about prior reactions to drugs, especially those containing polysorbate 80.
Have resuscitation equipment and trained personnel immediately available at the time of Vizamyl administration [see Contraindications (4)].
Errors may occur while using Vizamyl PET images to estimate brain neuritic plaque density [see Clinical Studies (14)].
Image interpretation is performed independently of the patient's clinical information. The use of clinical information in the interpretation of Vizamyl images has not been evaluated and may lead to errors. Extensive brain atrophy may limit the ability to distinguish grey and white matter on a Vizamyl scan [see Dosage and Administration (2.5)]. Motion artifacts may distort the image [see Dosage and Administration (2.3)].
Vizamyl scan results are indicative of the brain neuritic amyloid plaque content only at the time of image acquisition and a negative scan result does not preclude the development of brain amyloid in the future.
Vizamyl, similar to other radiopharmaceuticals, contributes to a patient's overall long-term cumulative radiation exposure. Long-term cumulative radiation exposure is associated with an increased risk of cancer. Ensure safe handling to protect patients and health care workers from unintentional radiation exposure [see Dosage and Administration (2.1)].
Clinical trials are conducted under widely varying conditions and adverse reaction rates observed in the clinical trials of Vizamyl cannot be directly compared to rates in the clinical trials of another drug and may not reflect the rates observed in clinical practice.
In clinical trials, 761 adults (367 men and 394 women, 91% Caucasian) with a mean age of 62 years (range 18-93 years) received Vizamyl. Most subjects (530, 70%) received a dose of 185 MBq (5 mCi).
One subject out of 761 administered Vizamyl experienced a serious hypersensitivity reaction with flushing, dyspnea and chest pressure within minutes following Vizamyl administration and recovered with treatment.
Most adverse reactions were mild to moderate in intensity and resolved spontaneously. The most commonly reported adverse reactions (occurring in at least 1% of subjects) in Vizamyl-treated subjects are shown in Table 2.
Adverse Reaction | N (percent of patients) |
---|---|
Flushing | 16 (2%) |
Increased blood pressure | 13 (2%) |
Headache | 10 (1%) |
Nausea | 8 (1%) |
Dizziness | 8 (1%) |
Pharmacodynamic drug-drug interaction studies have not been performed in patients to establish the extent, if any, to which concomitant medications may alter Vizamyl image results.
Within a clinical study of patients with a range of cognitive impairment, some patients were receiving the following medications: donepezil, galantamine, memantine, rivastigmine. Mean cortical Standardized Uptake Value (SUV) ratios did not differ between the patients taking or not taking these concomitant medications.
Risk Summary
There are no available data on Vizamyl use in pregnant women to evaluate for a drug-associated risk of major birth defects, miscarriage, or adverse maternal or fetal outcomes. All radiopharmaceuticals, including Vizamyl, have the potential to cause fetal harm depending on the stage of fetal development, and the magnitude of the radiopharmaceutical dose. If considering Vizamyl administration to a pregnant woman, inform the patient about the potential for adverse pregnancy outcomes based on the radiation dose from Vizamyl and the gestational timing of exposure.
The estimated background risk of major birth defects and miscarriage for the indicated population(s) is unknown. All pregnancies have a background risk of birth defect, loss, or other adverse outcomes. In the U.S. general population, the estimated background risk of major birth defects and miscarriage in clinically recognized pregnancies is 2 to 4% and 15 to 20%, respectively.
Risk Summary
There are no data on the presence of Flutemetamol (F 18) or metabolites in human milk or its effects on the breastfed infant or milk production. Exposure of Vizamyl to a breastfed infant can be minimized by temporary discontinuation of breastfeeding [see Clinical Considerations]. The developmental and health benefits of breastfeeding should be considered along with the mother's clinical need for Vizamyl and any potential adverse effects on the breastfed child from Vizamyl or from the underlying maternal condition.
The clinical consequence of overdose with Vizamyl has not been reported. It is unknown whether or not flutemetamol is dialyzable. The major risks of overdose relate predominantly to increased radiation exposure, with long-term risks for neoplasia. In case of overdose of radioactivity, hydration and frequent urination should be encouraged to minimize radiation exposure to the patient; care should be taken to avoid contamination from the radioactivity eliminated by the patient.
Vizamyl contains flutemetamol F 18, a molecular imaging agent that binds to β-amyloid aggregates, and is intended for use with PET imaging of the brain. Chemically, flutemetamol F 18, is described as 2-[3-[18F]fluoro-4-(methylamino) phenyl]-6-benzothiazolol. It has the molecular formula C14H1118FN2OS, molecular weight 273.32, and the following structural formula:
Vizamyl is a sterile, non-pyrogenic, clear, colorless to slightly yellow radioactive solution for intravenous injection. Each milliliter (mL) of the no-carrier added Vizamyl product solution contains 150 MBq (4.05 mCi) of flutemetamol F 18 at reference date and time and up to 2 micrograms of flutemetamol. Each mL of the Vizamyl solution also contains 70 microliters ethanol, 9.0 mg sodium chloride and 4.98 mg polysorbate 80 (w/v) in 0.014 M aqueous phosphate buffer. The pH of the solution is between 6.0 and 8.5.
Fluorine-18 (F 18) is a cyclotron-produced radionuclide that decays by positron emission (β+ decay, 96.7%) and orbital electron capture (3.3%) to stable oxygen-18 with a physical half-life of 109.8 minutes. The positron can undergo annihilation with an electron to produce two gamma rays; the energy of each gamma ray is 511 keV (Table 3).
Radiation | Energy (keV) | Abundance (%) |
---|---|---|
Gamma | 511 | 193.4 |
Positron | 249.8 | 96.7 |
The point source air-kerma rate constant for F-18 is 3.74E -17 Gy m2/(Bq s); this coefficient was formerly defined as the specific gamma-ray constant of 5.7 R/hr/mCi at 1 cm. The first half-value thickness of lead (Pb) for F 18 gamma rays is approximately 6 mm. The relative reduction of radiation emitted by F-18 that results from various thicknesses of lead shielding is shown in Table 4. The use of ~8 cm of Pb will decrease the radiation transmission (i.e., exposure) by a factor of about 10,000.
Shielding Thickness cm of lead (Pb) | Coefficient of Attenuation |
---|---|
0.6 | 0.5 |
2 | 0.1 |
4 | 0.01 |
6 | 0.001 |
8 | 0.0001 |
Flutemetamol F 18 binds to β-amyloid plaques in the brain and the F-18 isotope produces a positron signal that is detected by a PET scanner. In in vitro binding studies using postmortem human brain homogenates containing fibrillar β-amyloid, the dissociation constant (Kd) for flutemetamol was 6.7 nM.
Selectivity of [3H]flutemetamol binding in post-mortem human brain sections was demonstrated using autoradiography, silver-stained protein, and immunohistochemistry (monoclonal antibody to β-amyloid) correlation studies.
Following intravenous injection, flutemetamol F 18 diffuses across the human blood-brain barrier and produces a radioactivity signal detectable throughout the brain. Subsequently, cerebral perfusion decreases the brain flutemetamol F 18 content, with differential retention of the drug in cortical areas that contain β-amyloid aggregates compared to areas that lack the aggregates. The time-activity curves for flutemetamol F 18 in the brain of subjects with positive scans shows continual signal increases from time zero through 30 minutes post administration, with stable values thereafter up to at least 120 minutes post-injection. Differences in signal intensity between brain regions that specifically retain flutemetamol F 18 and brain regions with nonspecific retention of the drug form the basis of image interpretation methods [see Dosage and Administration (2.5)].
The test-retest distribution of flutemetamol F 18 was evaluated in 5 subjects with probable AD who underwent two administrations of flutemetamol F 18 (followed by PET scans) separated by a time period of 1 to 4 weeks. Images were reproducible when evaluated semi-quantitatively using an automated assessment of SUV in pre-specified cortical regions of brain.
Following intravenous injection of 185 MBq (5 mCi) of Vizamyl in humans, flutemetamol F 18 plasma concentrations declined by approximately 75% in the first 20 minutes post-injection, and by approximately 90% in the first 180 minutes. The F 18 in circulation during the 30-120 minutes imaging window in plasma was principally associated with flutemetamol metabolites. Excretion was approximately 37% renal (28-45%; n=6) and 52% hepatobiliary (40-65%; n=6).
Animal studies have not been performed to evaluate the carcinogenicity potential of flutemetamol. Flutemetamol was positive for mutagenicity in two in vitro assays: the bacterial reverse mutation assay (Ames test) and the mouse lymphoma assay.
Flutemetamol was negative for genotoxicity after in vivo exposure in rats to flutemetamol at the highest cumulative dose level tested, as measured in bone marrow micronucleus assays (157 and 27 microgram/kg/day for 2 and 14 days respectively) and an unscheduled DNA synthesis assay in rat hepatocytes (39 microgram/kg/day).
Vizamyl was evaluated in two clinical studies (Study One and Study Two) in adult subjects with a range of cognitive function, including some terminally ill patients who had agreed to participate in a post-mortem brain donation program. Both studies were single-arm and subjects underwent Vizamyl injection and scan. The images were interpreted by five independent readers masked to all clinical information; readers in Study Two were naïve to all forms of amyloid PET imaging. PET images were reviewed first without, and subsequently with, brain CT or MRI images.
Study One evaluated pre-mortem Vizamyl PET images from terminally ill patients and compared the results to postmortem truth standard assessments of cerebral cortical neuritic plaque density in patients who died during the study. Readers evaluated images using a clinically applicable binary image interpretation method (positive/negative) that involved evaluating regional Vizamyl brain uptake to yield a final overall image assessment that was compared to the truth standard. Before image interpretation, all readers underwent in-person tutoring on image interpretation. To determine the agreement between the in vivo Vizamyl image results and the post-mortem whole brain amyloid neuritic plaque density, Vizamyl results (negative/positive) were pre-specified to correspond with specific global histopathology plaque density scores, based upon a modification of the Consortium to Establish a Registry for Alzheimer's Disease (CERAD) criteria (Table 5), which use neuritic plaque counts as a necessary pathological feature of AD. Plaques were counted on microscope slides with modified Bielschowsky silver-stained tissue sections. The global brain neuritic plaque density score for each subject was determined by averaging across the scores (0-3) for five grey matter fields per slide and then across the six slides for each of eight regions; if any one region had a regional score of greater than 1.5, the subject's brain was classified as positive for amyloid.
Vizamyl Image Result | CERAD Classification (Score) | Neuritic Plaque Counts |
---|---|---|
Negative | None (0) | 0 |
Sparse (1) | 1 to 5 | |
Positive | Moderate (2) | 6 to 19 |
Frequent (3) | >20 |
In Study One, one hundred eighty patients were dosed with Vizamyl and 176 were imaged. The median patient age was 82 years (range 47 to 98 years) and 57% of the patients were female. By medical history 44 patients had no cognitive impairment, 135 had dementia, no patients had mild cognitive impairment (MCI), and one patient had memory loss of unspecified nature. Sixty-nine patients died during the study; 68 had cerebral cortical amyloid status determined (43 positive and 25 negative) and were included in the primary analysis. The time interval between the Vizamyl scan and death ranged from 0 to 13 months, with a median of 2.6 months, and was less than one year for 66 patients and between 12 to 13 months for 2 patients. At autopsy, the global brain neuritic plaque density category (CERAD classification as in Table 5) was available for 67/68 subjects: frequent (n = 19); moderate (n = 22); sparse (n = 14); and none (n = 12).
In Study Two, the effectiveness of an electronic training program for Vizamyl image orientation and interpretation was evaluated using Vizamyl PET images from across subjects with different cognitive abilities who had participated in earlier studies. Inter-reader reproducibility of image interpretation was assessed using images from subjects with a truth standard (68 patients who underwent an autopsy and 36 known or suspected normal pressure hydrocephalus patients with in vivo brain biopsy) and without a truth standard (28 cognitively normal volunteers 55 years or above, 80 patients with amnestic mild cognitive impairment (aMCI), 33 patients with probable AD (pAD)), and 31 young healthy volunteers. Additionally, intra-reader reproducibility was assessed from 29 images (10%). Among the 276 subjects, the median age was 72 years (range 20 to 95), 136 were females, and 251 were Caucasian.
Vizamyl performance characteristics for Study One and Study Two patients with an autopsy-based truth standard are shown in Table 6 and Table 7. Among patients who underwent autopsy (n=68; 43 positive and 25 negative based on histopathology), the median (and range) of correct read results, false negatives, and false positives were 59 (51, 61), 5 (3, 8), 3 (2, 14), respectively, for in-person training (Study One); and were 60 (55 to 61), 3 (3 to 6), 4 (2 to 10), respectively, for electronic media training (Study Two). Image reproducibility for various subject groups in Study Two is presented in Table 8. Inter-reader reproducibility analysis showed an overall Fleiss' kappa statistic of 0.83 (95% CI 0.79 to 0.86) which met the pre-specified success criterion (95% CI lower bound > 0.60). Intra-reader reproducibility analysis showed that, between the two readings for each of the 29 duplicate patient images, one of the five readers had complete agreement for all 29 images, two readers had discordant reads for a single image, and two readers had discordant reads for two images. Intra-reader reproducibility for a subgroup of 8 images from aMCI patients showed that all five readers had complete agreement for all duplicate images.
Test Performance | In-Person Training (Study One) | Electronic Media Training (Study Two) |
|
---|---|---|---|
Sensitivity (%) | Median | 88 | 93 |
Range among the 5 readers | 81 - 93 | 86 - 93 | |
Specificity (%) | Median | 88 | 84 |
Range among the 5 readers | 44 - 92 | 60 - 92 |
In-Person Training (Study One) | Electronic Media Training (Study Two) |
||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Reader | Reader | ||||||||||
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | ||
|
|||||||||||
All scans with autopsies (n = 68*) | Correct | 57 | 60 | 51 | 59 | 61 | 58 | 61 | 61 | 55 | 60 |
False Negative | 8 | 5 | 3 | 3 | 5 | 3 | 3 | 4 | 3 | 6 | |
False Positive | 3 | 2 | 14 | 5 | 2 | 7 | 4 | 3 | 10 | 2 |
Subject Group by Cognitive and Truth Standard (TS) | Positive Scans N* | Kappa (95% CI) | Percent of Scans with Inter-reader Agreement | ||
---|---|---|---|---|---|
3 of 5 readers agreed | 4 of 5 readers agreed | 5 of 5 readers agreed | |||
pAD: probable AD; aMCI: amnestic MCI; Elderly: 55 years or above | |||||
|
|||||
All 276 subjects | 139 | 0.83 (0.79, 0.86) | 5 | 14 | 81 |
All subjects with a TS, n=104 (68 autopsy; 36 biopsy) | 58 | 0.74 (0.68, 0.80) | 6 | 24 | 70 |
All subjects without a TS, n = 172 | 76 | 0.88 (0.83, 0.92) | 5 | 8 | 87 |
pAD, n = 63 (30 with TS†; 33 no TS) | 47 | 0.88 (0.80, 0.96) | 3 | 6 | 90 |
aMCI, n = 80 (0 with TS) | 45 | 0.89 (0.82, 0.96) | 4 | 7 | 89 |
Elderly cognitively normal without TS, n = 28 | 2 | 0.46 (0.34, 0.57) | 4 | 14 | 82 |
Cognitively normal with TS‡, (n=21) | 10 | 0.64 (0.5, 0.77) | 5 | 38 | 57 |
Other (non-AD) dementia with TS, n=53§ | 27 | 0.71 (0.63, 0.80) | 8 | 25 | 68 |
Vizamyl is supplied in a 30-mL multi-dose glass vial with 1-30 mL fill volume. Each vial is enclosed in an appropriate radiation shield. The total concentration is 150 MBq/mL (4.05 mCi/mL) of flutemetamol F 18 at reference date and time.
30-mL sterile multi-dose vial with variable fill volume: NDC: 17156-067-30
Instruct patients to inform their healthcare provider if they:
Instruct patients to increase their level of hydration before and after receiving Vizamyl (Flutemetamol F 18 Injection) and to void frequently for the first 24 hours following Vizamyl administration.
Manufactured for GE Healthcare, Medi-Physics, Inc., Arlington Heights, IL 60004 U.S.A.
Vizamyl is a trademark of General Electric Company or one of its subsidiaries.
GE and the GE Monogram are trademarks of General Electric Company.
© 2020 General Electric Company – All rights reserved.
43-1067D
NDC: 17156-067-30
Vizamyl™
Flutemetamol F 18 Injection
30 mL Multiple-Dose Vial
Sterile Non-Pyrogenic
Rx ONLY
CAUTION
RADIOACTIVE
MATERIAL
_______ MBq (________ mCi) in ________ mL at ______:______ on _______________
Batch No.: _______________________ Date: ___________________ Vial No.: _______
Diagnostic - For Intravenous Use Only.
Contents per mL of Vizamyl (Flutemetamol F 18 Injection) solution at
reference time: ____________________ , up to 2 micrograms of flutemetamol and
150 MBq (4.05 mCi) [18F]flutemetamol, 7% ethanol (v/v), 0.9% sodium chloride
(w/v) and 0.5% polysorbate 80 (w/v) in 0.014 M phosphate buffer.
Store at 2°C to 30°C (35.6°F to 86°F).
Expires at ______:______ on _________________
Manufactured by Cardinal Health 414, LLC, Dublin, OH 43017
for GE Healthcare, Medi-Physics Inc., Arlington Heights, IL 60004, U.S.A.
41-3067
VIZAMYL
flutemetamol f-18 solution |
||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||
|
Labeler - Medi-Physics, Inc. dba GE Healthcare (095263729) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health 414, LLC | 165486861 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health 414, LLC | 833114734 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Precision Nuclear, LLC | 879283633 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health, Inc. (East Lansing, MI) | 963972166 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health 414, LLC | 964767656 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health 414, LLC | 964767771 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health 414, LLC | 964768126 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health 414, LLC | 964768282 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health 414, LLC | 964768340 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Cardinal Health 418, Inc. (Aurora, CO) | 149029253 | ANALYSIS(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
GE Healthcare AS (Oslo, Norway) | 515048908 | MANUFACTURE(17156-067) , ANALYSIS(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
The University of Utah,DBA CyclotronRadiochemistry Lab Huntsman Cancer Institute (Salt Lake City, UT) | 018432646 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Essential Isotopes, LLC. (Colombia, MO) | 010753961 | MANUFACTURE(17156-067) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Medi-Physics, Inc. dba GE Healthcare | 095263729 | ANALYSIS(17156-067) |
Mark Image Registration | Serial | Company Trademark Application Date |
---|---|
VIZAMYL 86968004 not registered Dead/Abandoned |
GE Healthcare Limited 2016-04-07 |
VIZAMYL 85721037 not registered Dead/Abandoned |
GE Healthcare Limited 2012-09-05 |
VIZAMYL 85328526 4659887 Live/Registered |
GE Healthcare Limited 2011-05-24 |