MAUDE MDR 2750239

MDR report key
2750239
Report number
9617544-2012-00393
Event key
0
Event type
3
Date of event
2012-08-17
Date received
2012-09-14
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
0

Manufacturer Contact#

Contact
SONIA POUSTIS
Address
2 PEARL COURT ALLENDALE NJ 07450 US
Phone
201-201-2010
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1REFLEX HYBRID REVISION DRIVER DRAW ROD SCREW EXTRAINSTRUMENTSTRYKER SPINE BORDEAUXLXHNA48511906B109528Y Y

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12012-09-140

Event Narratives#

D

Patient 1

HOLD - SDAIT WAS REPORTED THAT, "REFLEX HYBRID REVISION DRIVER DRAW ROD TIP BROKE OFF DURING REMOVAL OF SCREW. SCREW HAD ALREADY CLEARED THE LOCKING RING. TIP WAS VISIBLE INSIDE. REMOVED SCREW AND DISPOSED OF. USED BACK UP REVISION DRIVER TO PROCEED WITH NO DELAY IN THE PROCEDURE AND NO LOST OPERATING ROOM TIME."

N

Patient 1

METHOD: VISUAL INSPECTION, MANUFACTURING RECORD REVIEW, COMPLAINT HISTORY ANALYSIS AND RISK ASSESSMENT. RESULTS: AFTER VISUAL INSPECTION, THE TIP WAS CONFIRMED TO BE BROKEN. MANUFACTURING RECORD REVIEW: NO ANOMALIES THAT COULD BE ASSOCIATED WITH THE BREAKAGE WERE REPORTED DURING THE MANUFACTURING. COMPLAINT HISTORY ANALYSIS: 90 PREVIOUS RECORDS HAVE BEEN RECEIVED INVOLVING 93 UNITS. INVESTIGATIONS INTO PAST COMPLAINTS CONCLUDED THAT THE FAILURES WERE THE RESULT OF USER-ERROR I.E. OVER-ANGULATION. RISK ASSESSMENT: THERE WERE NO REPORTS OF ANY ADVERSE CONSEQUENCES TO THE PATIENT OR OF ANY DELAYS IN SURGERY. THE REFLEX-HYBRID SCREW EXTRACTOR INNER SHAFT IS MADE OF IMPLANT GRADE BIOCOMPATIBLE STAINLESS STEEL TO MITIGATE THE RISK OF BROKEN TIPS REMAINING INSIDE THE PATIENT SHOULD THE DEVICE BREAK DURING SURGERY. CONCLUSION: THE INSTRUMENT FAILURE IS BELIEVED TO BE THE RESULT OF USER-ERROR. THE SURGICAL TECHNIQUE STATES THAT PIVOTING OR ANGULATION OF THE INSTRUMENT MUST BE AVOIDED AS THIS CAN CAUSE BENDING OR BREAKING OF THE INNER SHAFT.