MAUDE MDR 2980991

MDR report key
2980991
Report number
1028232-2013-00448
Event key
0
Event type
3
Date of event
2012-09-24
Date received
2013-02-19
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
2
Health professional
3
Initial report to FDA
3
Event location
0

Manufacturer Contact#

Address
6024 JEAN ROAD LAKE OSWEGO OR 97035 US
Phone
877-877-8772
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1SETROX S 45PACER LEADBIOTRONIK SE & CO. KGJXK350973SEE MODEL NO.R N

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12013-02-1901. H

Event Narratives#

D

Patient 1

THIS SYSTEM WAS REMOVED DUE TO (B)(6) ENDOCARDITIS ON (B)(6) 2012. THERE IS NO INDICATION THE SYSTEM WAS REPLACED. SHOULD ADDITIONAL INFO BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

N

Patient 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE WERE RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.