N
Patient 1
THE ARGON PLASMA COAGULATOR (APC)/ELECTROSURGICAL UNIT (ESU) SYSTEM WAS RETURNED AND THOROUGHLY EVALUATED. THE FINDINGS WERE AS FOLLOWS: CHRONOLOGICAL LOG REVIEW: THE LOG ON THE DAY OF THE REPORTED EVENT WAS REVIEWED. NO SYSTEM ISSUES WERE RECORDED. APC/ESU INSPECTION AND TESTING: A TECHNICAL SAFETY CHECK WAS PERFORMED ON EACH UNIT. THIS INCLUDED AN ELECTRICAL SAFETY CHECK, A FUNCTION CHECK OF EACH OF THE EQUIPMENT'S FEATURES, AND A POWER OUTPUT CHECK. ALSO, THE GAS FLOW RATES WERE MEASURED AND FOUND TO BE WITHIN THEIR ACCEPTABLE RANGES FOR THE APC. ALL FEATURES WERE/ARE FUNCTIONING PROPERLY WITHIN SPECIFICATIONS ON BOTH DEVICES. IN ADDITION, NO ANOMALIES WERE FOUND IN THE DEVICE HISTORY RECORDS (DHRS) FOR THE APC AND ESU. IN CONCLUSION, NO ERBE EQUIPMENT PROBLEM WAS FOUND THAT WOULD HAVE CAUSED OR ATTRIBUTED TO THE EVENT. NO DETERMINATION COULD BE MADE AS TO THE CAUSE OF THE EVENT. THE ACCOUNT IS BEING MADE AWARE OF THE FINDINGS. TO FURTHER ADDRESS THE SITUATION, ADDITIONAL CASE WORK IS BEING PLANNED WITH THE DOCTOR. NO TRENDS HAVE BEEN IDENTIFIED WITH THIS INCIDENT. ERBE USA, INC. IS NOW CLOSING THE FILE ON THIS EVENT.