PARKER BATH AL14110-EU

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2019-01-18 for PARKER BATH AL14110-EU manufactured by Arjo Hospital Equipment Ab.

Event Text Entries

[134572175] (b)(4). Arjo representative evaluated the device in question. The gas strut head was loose because of the absence of a locknut. The gas strut head was disassembled and the door was completely free, so it was closing heavy. The bathtub was not removed from use by the customer facility despite the problem. The defective part was replaced by the arjo representative. Additional information will be provided within the next report.
Patient Sequence No: 1, Text Type: N, H10


[134572176] It was reported that during inspection requested by the customer facility, the arjo qualified representative detected the defect of the bath's door gas strut. The malfunction resulted in the door unable to remain in required position and closing fast. The bathtub was used by the customer despite the malfunction. No information regarding an injury occurrence was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007420694-2019-00018
MDR Report Key8259961
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2019-01-18
Date of Report2019-03-08
Date Facility Aware2018-12-20
Report Date2019-03-08
Date Reported to FDA2019-03-08
Date Reported to Mfgr2019-03-08
Date Mfgr Received2018-12-20
Device Manufacturer Date2005-11-14
Date Added to Maude2019-01-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KINGA STOLINSKA
Manufacturer StreetKS. WAWRZYNIAKA 2
Manufacturer CityKOMORNIKI, 62-052
Manufacturer CountryPL
Manufacturer Postal62-052
Manufacturer G1ARJO HOSPITAL EQUIPMENT AB
Manufacturer StreetVERKSTADSVAGEN 5
Manufacturer CityESLOV, 24121
Manufacturer CountrySW
Manufacturer Postal Code24121
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePARKER BATH
Generic NameBATH, HYDRO-MASSAGE
Product CodeILJ
Date Received2019-01-18
Model NumberAL14110-EU
Device AvailabilityY
Device Age13 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerARJO HOSPITAL EQUIPMENT AB
Manufacturer AddressVERKSTADSVAGEN 5 ESLOV, 24121 SW 24121


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-01-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.