MAUDE MDR 9392183

MDR report key
9392183
Report number
2032227-2019-125345
Event key
0
Event type
3
Date of event
2019-09-16
Date received
2019-11-30
Adverse event
0
Product problem
1
Patients in event
0
Reporter occupation
0
Health professional
0
Initial report to FDA
0
Event location
0

Manufacturer Contact#

Contact
GERWIN DE GRAAFF
Address
CEIBA NORTE IND. PARK #50 ROAD JUNCOS 00777
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1PUMP MMT-1780KPK 670G PATHWAY BLACK MGARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROLMEDTRONIC PUERTO RICO OPERATIONS CO.OZPMMT-1780KPKMMT-1780KPKHG30NCXR R
1017937PUMP MMT-1780KPK 670G PATHWAY BLACK MGARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROLMEDTRONIC PUERTO RICO OPERATIONS CO.OZPMMT-1780KPKMMT-1780KPKHG30NCXRR

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12019-11-300

Event Narratives#

N

Patient 1

(B)(4). PUMP RECEIVED WITH CRACKED CASE BATTERY TUBE, CRACKED CASE CORNER OF BELT CLIPS RAILS, DETACHED/MISSING RESERVOIR RETAINER RING, BROKEN RESERVOIR TUBE LIP AND MISSING RESERVOIR TUBE LIP O-RING NOTED. THE TEST RESERVOIR DOES NOT STAY LOCKED IN PLACE DUE TO DETACHED/MISSING RETAINER. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO DETACHED/MISSING RETAINER.

D

Patient 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT LEFT OF BELT CLIP GOING DOWN OF INSULIN PUMP WAS CRACKED. THE CUSTOMER?S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. CUSTOMER ALSO STATED RESERVOIR WAS DAMAGED AND RESERVOIR LOCKED IN PLACE WHILE INSERTED. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN PER HEALTH CARE PROFESSIONAL. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

N

Patient 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. (B)(4). MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.