O2 BI70002000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-26 for O2 BI70002000 manufactured by Sanmina -sci Systems.

Event Text Entries

[186731952] The system was serviced in the field. There was a loss of calibration files and a loss of communication between the mvs and ias computers. The field service engineer (fse) checked, cleaned and re-seated all ethernet cables and connections. The fse reloaded the mvs configuration files. He tested ethernet communication, verify complete boot, system operation and re-connectivity with umbilical dis-connection and re-connection. He completed power cycles and performed a system test. The system performed to specification. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[186731953] Medtronic received information regarding an imaging system used outside of procedure. It was reported that the image acquisition station (ias) was not connecting to the mobile viewing station (mvs) and had an error "unable to connect to gain calibration file path". The radiologic technologist (rt) tried multiple reboots and the issue did not resolve. There was no patient present at the time of the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006544299-2020-00103
MDR Report Key9883691
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-24
Date Mfgr Received2020-03-24
Device Manufacturer Date2018-10-08
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1SANMINA -SCI SYSTEMS
Manufacturer StreetKM 15.5 NO. 29, PLANT 06
Manufacturer CityCARR. CHAPALA-GUADALAJARA,JAL 45640
Manufacturer CountryMX
Manufacturer Postal Code45640
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameO2
Generic NameINTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Product CodeOWB
Date Received2020-03-26
Model NumberBI70002000
Catalog NumberBI70002000
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSANMINA -SCI SYSTEMS
Manufacturer AddressKM 15.5 NO. 29, PLANT 06 CARR. CHAPALA-GUADALAJARA,JAL 45640 MX 45640


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.