MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-30 for O-ARM SYS O2 BI70002000 manufactured by Medtronic Navigation, Inc (littleton).
[186183610]
A medtronic representative went to the site to perform a system checkout. They confirmed generator error 11, and no voltage on the caps. They also found the charger assembly to not be allowing voltage to be output to the generator. The charger assembly was replaced and the system now functions as intended. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[186183611]
Medtronic received information regarding an imaging system used outside of procedure. It was reported that upon boot up the image acquisition station (ias) was beeping, indicating a generator error. There was no patient present.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004785967-2020-00426 |
MDR Report Key | 9900878 |
Report Source | COMPANY REPRESENTATIVE |
Date Received | 2020-03-30 |
Date of Report | 2020-03-30 |
Date of Event | 2020-03-16 |
Date Mfgr Received | 2020-03-16 |
Device Manufacturer Date | 2016-09-02 |
Date Added to Maude | 2020-03-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | STACY RUEMPING |
Manufacturer Street | 7000 CENTRAL AVENUE NE RCW215 |
Manufacturer City | MINNEAPOLIS MN 55432 |
Manufacturer Country | US |
Manufacturer Postal | 55432 |
Manufacturer Phone | 7635260594 |
Manufacturer G1 | MEDTRONIC NAVIGATION, INC (LITTLETON) |
Manufacturer Street | 300 FOSTER ST |
Manufacturer City | LITTLETON MA 01460 |
Manufacturer Country | US |
Manufacturer Postal Code | 01460 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | O-ARM SYS O2 |
Generic Name | INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM |
Product Code | OWB |
Date Received | 2020-03-30 |
Returned To Mfg | 2020-03-18 |
Model Number | BI70002000 |
Catalog Number | BI70002000 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC NAVIGATION, INC (LITTLETON) |
Manufacturer Address | 300 FOSTER ST LITTLETON MA 01460 US 01460 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-30 |