O-ARM O2 IMAGING SYSTEM BI70002000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-31 for O-ARM O2 IMAGING SYSTEM BI70002000 manufactured by Medtronic Navigation, Inc (littleton).

Event Text Entries

[185781338] A medtronic representative went to the site to perform a system check out and they found that the real problem was that system would not make an x-ray and needed a system reboot. The system reboot corrected all problems, and the system functioned as intended. The field service engineer also performed a user detector calibration as a precaution. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185781339] Medtronic received information regarding an imaging system used during a sacroiliac and thoracolumbar procedure. It was reported that after the initial spin the system went back to initializing. It was further stated that after initial boot up it was stuck in initializing, but that a reboot had resolved it. The site rebooted again with the image acquisition system (ias) and mobile viewing station (mvs) disconnected, connecting them again after boot up. This appeared to resolve the issue. There was no reported delay to the procedure. There was no impact to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004785967-2020-00430
MDR Report Key9904096
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-11
Date Mfgr Received2020-03-11
Device Manufacturer Date2018-01-25
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Street300 FOSTER ST
Manufacturer CityLITTLETON MA 01460
Manufacturer CountryUS
Manufacturer Postal Code01460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameO-ARM O2 IMAGING SYSTEM
Generic NameINTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Product CodeOWB
Date Received2020-03-31
Model NumberBI70002000
Catalog NumberBI70002000
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Address300 FOSTER ST LITTLETON MA 01460 US 01460


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.