| Seq | Brand | Generic | Manufacturer | Product code | Model | Catalog | Lot | PMA | 510(k) | Implant | Evaluated | Availability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | DEXCOM, INC. | QBJ | 9445-02 | STT-OR-001 | 5261574 | R | N | |||
| 2193327 | DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | DEXCOM, INC. | QBJ | 9445-02 | STT-OR-001 | 5261574 | R |
| Sequence | Received | Treatment | Outcome |
|---|---|---|---|
| 1 | 2020-03-31 | 0 |
Patient 1
(B)(4).
Patient 1
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient 1
(B)(4).
Patient 1
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND PASSED. PAIRING TEST WAS PERFORMED AND PASSED. A REVIEW OF THE BIN FILE WAS PERFORMED AND THE COMPLAINT STATES THAT A TRANSMITTER FAILED ERROR OCCURRED WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient 1
(B)(4).
Patient 1
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND PASSED. PAIRING TEST WAS PERFORMED AND PASSED. A REVIEW OF THE BIN FILE WAS PERFORMED AND THE COMPLAINT STATES THAT A TRANSMITTER FAILED ERROR OCCURRED WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.