MAUDE MDR 9908483

MDR report key
9908483
Report number
3004753838-2020-31131
Event key
0
Event type
3
Date of event
2020-03-14
Date received
2020-03-31
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
MS ASHLEY SPOTO
Address
6340 SEQUENCE DR. SAN DIEGO CA 92121 US
Phone
858-858-8582
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEMCONTINUOUS GLUCOSE MONITORDEXCOM, INC.QBJ9500-45STS-OR-001NIR N

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12020-03-310

Event Narratives#

N

Patient 1

(B)(4).

D

Patient 1

IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM,?WHICH IS OFF-LABEL USAGE OF THE DEVICE,?ON?(B)(6) 2020. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE B ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.