MAUDE MDR 9908928

MDR report key
9908928
Report number
3004209178-2020-92492
Event key
0
Event type
3
Date of event
2019-11-07
Date received
2020-03-31
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
GERWIN DE GRAAFF
Address
CEIBA NORTE IND. PARK #50 ROAD JUNCOS 00777 *
Phone
818-818-8185
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1PUMP MMT-1714K 630G BLACK MMOL CANADAARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPENDMEDTRONIC PUERTO RICO OPERATIONS CO.OZOMMT-1714KMMT-1714KHG29CXPR R

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12020-03-310

Event Narratives#

N

Patient 1

(B)(4). THE PUMP WAS RECEIVED WITH A BROKEN RESERVOIR TUBE LIP, MISSING RETAINER AND MISSING RESERVOIR TUBE O-RING. UNIT P-CAP / TEST RESERVOIR DOES NOT LOCK IN PLACE PROPERLY AND UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO THE MISSING RETAINER AND BROKEN RESERVOIR TUBE LIP.

D

Patient 1

INFORMATION RECEIVED BY THE MEDTRONIC INDICATED THAT THE INSULIN PUMP?S RESERVOIR LIP RING WAS COMING OFF HOWEVER RESERVOIR WAS ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.