PMA P010025S017

Device
SELENIA DIMENSIONS 2D MOBILE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Applicant
Hologic, Inc.
PMA number
P010025
Supplement
S017
Product code
MUE
Decision date
2009-12-23
Generic name
Full field digital, system, x-ray, mammographic
Approval order statement
APPROVAL FOR THE SELENIA DIMENSION 2D FFDM TO BE USED IN A MOBILEENVIRONMENT WITH THE FOLLOWING DESIGN CHANGES:1) ACQUISITION STATION:A) THE ACQUISITION STATION MONITOR WILL BE MOUNTED ON A FIXED MOUNT MECHANISM THAT ALLOWS POSITIONING OF THE MONITOR VIA A SPRING LOADED LOCKING DEVICE. THE MOBILE SYSTEM MONITOR CAN BE FIXED MOUNTED ON THE LEFT,RIGHT OR CENTER POSITION.B) THE KEYBOARD TRAY HAS BEEN SECURED WITH A SPRING LOADED LATCHING MECHANISM TO PREVENT MOVEMENT DURING TRANSPORT.C) THE BAR CODE SCANNER CAN BE MOUNTED EITHER ON THE LEFT OR RIGHT SIDE OF THE MONITOR AS NEEDED IN THE MOBILE APPLICATION.D) THE X-RAY SHIELD IS REPLACED WITH NON ATTENUATING, COSMETIC CLEAR PLASTIC (AS COVER GAP FILLERS). 2) GANTRY:A) THE VERTICAL TRAVEL ASSEMBLY (VTA) HAS BEEN CONFIGURED WITH AN ELECTROMAGNETICBRAKE TO PREVENT UNWANTED TRAVEL DURING TRANSPORT. B) A MOUNTING FEATURE HAS BEEN ADDED TO THE TOP SECTION OF THE GANTRY FRAME FORSUPPLEMENTAL SUPPORT IN A COACH/VAN.C) THE UPPER MOVEMENT LIMIT OF THE C-ARM IS NOW CONFIGURABLE FROM 6' 10" TO 7' 10" TO ALLOW FOR LOWER CEILINGS IN A COACH/VAN.3) INSTRUCTIONS FOR USE AND QC MANUAL:A MOBILE ADDENDUM INCLUDES CURRENT USER MANUAL WITH ADDED WARNINGS AND INSTRUCTIONSREGARDING MOBILE USE. IN ADDITION, THE MOBILE ADDENDUM INCLUDES INSTRUCTIONS TO THE USER ON THE QC TESTS THAT ARE REQUIRED TO BE PERFORMED PRIOR TO CLINICAL USE AFTER MOVEMENT.

Current openFDA PMA Record#

Device
SELENIA DIMENSIONS 2D MOBILE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Applicant
Hologic, Inc.
PMA number
P010025
Supplement
S017
Product code
MUE
Generic name
Full field digital, system, x-ray, mammographic
Decision date
2009-12-23
Decision code
APPR
Date received
2009-11-05
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Approval order statement
APPROVAL FOR THE SELENIA DIMENSION 2D FFDM TO BE USED IN A MOBILEENVIRONMENT WITH THE FOLLOWING DESIGN CHANGES:1) ACQUISITION STATION:A) THE ACQUISITION STATION MONITOR WILL BE MOUNTED ON A FIXED MOUNT MECHANISM THAT ALLOWS POSITIONING OF THE MONITOR VIA A SPRING LOADED LOCKING DEVICE. THE MOBILE SYSTEM MONITOR CAN BE FIXED MOUNTED ON THE LEFT,RIGHT OR CENTER POSITION.B) THE KEYBOARD TRAY HAS BEEN SECURED WITH A SPRING LOADED LATCHING MECHANISM TO PREVENT MOVEMENT DURING TRANSPORT.C) THE BAR CODE SCANNER CAN BE MOUNTED EITHER ON THE LEFT OR RIGHT SIDE OF THE MONITOR AS NEEDED IN THE MOBILE APPLICATION.D) THE X-RAY SHIELD IS REPLACED WITH NON ATTENUATING, COSMETIC CLEAR PLASTIC (AS COVER GAP FILLERS). 2) GANTRY:A) THE VERTICAL TRAVEL ASSEMBLY (VTA) HAS BEEN CONFIGURED WITH AN ELECTROMAGNETICBRAKE TO PREVENT UNWANTED TRAVEL DURING TRANSPORT. B) A MOUNTING FEATURE HAS BEEN ADDED TO THE TOP SECTION OF THE GANTRY FRAME FORSUPPLEMENTAL SUPPORT IN A COACH/VAN.C) THE UPPER MOVEMENT LIMIT OF THE C-ARM IS NOW CONFIGURABLE FROM 6' 10" TO 7' 10" TO ALLOW FOR LOWER CEILINGS IN A COACH/VAN.3) INSTRUCTIONS FOR USE AND QC MANUAL:A MOBILE ADDENDUM INCLUDES CURRENT USER MANUAL WITH ADDED WARNINGS AND INSTRUCTIONSREGARDING MOBILE USE. IN ADDITION, THE MOBILE ADDENDUM INCLUDES INSTRUCTIONS TO THE USER ON THE QC TESTS THAT ARE REQUIRED TO BE PERFORMED PRIOR TO CLINICAL USE AFTER MOVEMENT.