- Device
- Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG)
- Applicant
- ABBOTT MEDICAL
- PMA number
- P010032
- Supplement
- S178
- Product code
- LGW
- Generic name
- Stimulator, spinal-cord, totally implanted for pain relief
- Decision date
- 2021-09-27
- Decision code
- APPR
- Date received
- 2021-06-29
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Approval order statement
- Approval to 1) Adding an existing Abbott qualified supplier for diode components for 3 types of diodes. 2) Replacing tantalum decoupling capacitors with ceramic capacitors sourced from an existing Abbott qualified supplier; and 3) Reducing the number of tantalum output capacitors on the EPPIC charge pump.