PMA P030054S370

Device
Quadra Assura, Unify, Unify Assura, and Unify Quadra CRT-D Devices
Applicant
ABBOTT MEDICAL
PMA number
P030054
Supplement
S370
Product code
NIK
Decision date
2019-09-04
Classification
Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (crt-d)
Generic name
Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)
Approval order statement
Replace bond tabs and epoxy with aluminum wire ultrasonically bonded directly to the substrate on high voltage hybrid assemblies.

Current openFDA PMA Record#

Device
Quadra Assura, Unify, Unify Assura, and Unify Quadra CRT-D Devices
Applicant
ABBOTT MEDICAL
PMA number
P030054
Supplement
S370
Product code
NIK
Generic name
Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)
Decision date
2019-09-04
Decision code
OK30
Date received
2019-08-07
Supplement type
30-Day Notice
Supplement reason
Process Change - Manufacturer/Sterilizer/Packager/Supplier
Approval order statement