PMA P050027S022

Device
KARL STORZ Photodynamic D-Light C PDD System
Applicant
KARL STORZ Endoscopy-America, Inc.
Product code
OAY
Decision date
2020-08-17
Generic name
Light source system, diagnostic endoscopic
Approval order statement
Approval for increase in the tolerance of the countersink angle of the outer shell of the Grasping Mechanism of the Tricam camera heads from 82 +/-I degree to 82 +7/-3 degrees.

Current openFDA PMA Record#

Device
KARL STORZ Photodynamic D-Light C PDD System
Applicant
KARL STORZ Endoscopy-America, Inc.
PMA number
P050027
Supplement
S022
Product code
OAY
Generic name
Light source system, diagnostic endoscopic
Decision date
2020-08-17
Decision code
APPR
Date received
2020-05-08
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Approval order statement