- Device
- NeuroPace RNS System
- Applicant
- NeuroPace, Inc.
- PMA number
- P100026
- Supplement
- S075
- Product code
- PFN
- Generic name
- Implanted brain stimulator for epilepsy
- Decision date
- 2019-12-17
- Decision code
- APPR
- Date received
- 2019-09-30
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Approval order statement
- Approval for 1) a firmware update (RAM 7.101) to the RNS Neurostimulator (model RNS-320) to disable Long Episode counting during telemetry; and 2) an adjustment to the crystal oscillator trim value (from 00h to 64h) on the neurostimulator to mitigate the possibility of the device unnecessarily transitioning into passive mode.