- Device
- DEXCOM G5 MOBILE CONTIUOUS GLUCOSE MONITORING SYSTEM
- Applicant
- Dexcom, Inc.
- PMA number
- P120005
- Supplement
- S049
- Product code
- MDS
- Generic name
- Sensor, glucose, invasive
- Decision date
- 2016-09-16
- Decision code
- APPR
- Date received
- 2016-06-21
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Approval order statement
- Approval for updates to the G5 Mobile iOS App Software and G5 Mobile Transmitter Firmware for Dexcom’s G5 Mobile Continuous Glucose Monitoring System. The App is being modified to: backfill data from the CGM if there are data gaps within the last 3 hours; include mute override settings, which allow users to hear alerts critical to safe and effective use of the device even when their smart device is muted or turned to a very low volume; and to correct software anomalies. The Transmitter Firmware is being modified to: transmit data for backfill in the App; improve efficiency of battery usage on the transmitter; expand transmitter database logging to include additional information; support internal engineering efforts; and to correct software anomalies.