PMA P840001S628

Device
Master Restore™; Itrel™; Synergy™; Intellis™; Vanta™ and Inceptiv™ Spinal Cord Stimulation Systems and Pisces™; Specify™
Applicant
Medtronic Neuromodulation
PMA number
P840001
Supplement
S628
Product code
LGW
Decision date
2026-06-05
Classification
Neurology
Generic name
Stimulator, spinal-cord, totally implanted for pain relief
Approval order statement
for expansion of incoming electrolyte salt acceptance limits. Specifically, to expand LiAsF6 electrolyte salt moisture content acceptance limit from: = 100 parts per million (ppm) to: = 200 ppm and expand LiAsF6 electrolyte salt insoluble acceptance limit from: = 20 ppm to: = 200 ppm

Current openFDA PMA Record#

Device
Master Restore™; Itrel™; Synergy™; Intellis™; Vanta™ and Inceptiv™ Spinal Cord Stimulation Systems and Pisces™; Specify™
Applicant
Medtronic Neuromodulation
PMA number
P840001
Supplement
S628
Product code
LGW
Generic name
Stimulator, spinal-cord, totally implanted for pain relief
Decision date
2026-06-05
Decision code
OK30
Date received
2026-05-08
Supplement type
30-Day Notice
Supplement reason
Process Change - Manufacturer/Sterilizer/Packager/Supplier
Approval order statement
for expansion of incoming electrolyte salt acceptance limits. Specifically, to expand LiAsF6 electrolyte salt moisture content acceptance limit from: = 100 parts per million (ppm) to: = 200 ppm and expand LiAsF6 electrolyte salt insoluble acceptance limit from: = 20 ppm to: = 200 ppm