- Device
- FACT(TM)AND ARC BALLOON CORNARY DILATATION CATHETERS
- Applicant
- Abbott Vascular Devices
- PMA number
- P910031
- Supplement
- S011
- Product code
- LOX
- Generic name
- Catheters, transluminal coronary angioplasty, percutaneous
- Decision date
- 1997-04-09
- Decision code
- APPR
- Date received
- 1996-08-26
- Supplement type
- Normal 180 Day Track
- Supplement reason
- Change Design/Components/Specifications/Material
- Approval order statement
- APPROVAL FOR THE FOLLOWING: 1) A CHANGE IN THE BALLOON DESIGN FROM A DOUBLE LAYER OF POLYETHYLENE WITH A POLYETHYLENE TEREPHALATE CENTER REGION TO A SINGLE LAYER OF POLYETHYLENE TEREPHALATE; 2) A CHANGE IN CATHETER SHAFT MATERIAL FROM POLYETHYLENE TO NYLON; 3) A CHANGE IN THE INTERNAL SHAFT DESIGN FROM A TWO LUMEN SIDE BY SIDE DESIGN TO A TWO LUMEN COAXIAL DESIGN; 4) A CHANGE IN THE BALLOON/SHAFT BONDING METHOD FROM HEAT SEALING TO A URETHANE ADHESIVE; 5) A CHANGE IN THE BALLOON LENGTHS FROM 25 MM TO 22 MM AND FROM 20 MM TO 18 MM; AND 6) A CHANGE IN THE COLOR OF THE CATHETER SHAFT FROM BLUE TO TEAL.