This medical device record is a pma supplement. A supplement may have changed the device description/function or indication from that approved in the original pma. Be sure to look at the original pma record for more information.
Approval for four edits in engineering drawing of "beam splitter, energy detector'' and change of the material in the beam splitter component of the energy detector from corning fused silica code 7940 to corning 7980 hpfs.
Device | STAR S4 IR Excimer Laser System, WaveScan WaveFront System, and iDesign Advanced WaveScan Studio System |
Classification Name | Excimer Laser System |
Generic Name | Excimer Laser System |
Applicant | AMO Manufacturing USA, LLC |
Date Received | 2018-12-31 |
Decision Date | 2019-04-12 |
PMA | P910062 |
Supplement | S008 |
Product Code | LZS |
Advisory Committee | Ophthalmic |
Supplement Type | Real-time Process |
Supplement Reason | Change Design/components/specifications/material |
Expedited Review | No |
Combination Product | No |
Applicant Address | AMO Manufacturing USA, LLC 510 Cottonwood Drive milpitas, CA 95035 |
Supplement Number | Date | Supplement Type |
---|---|---|
P910062 | Original Filing | |
S008 | 2018-12-31 | Real-time Process |
S007 | 2014-07-07 | Real-time Process |
S006 | 2014-05-19 | 30-day Notice |
S005 | ||
S004 | 2014-05-01 | 30-day Notice |
S003 | ||
S002 | ||
S001 | 2000-04-12 | Real-time Process |