PMA P950005S041

Device
BI-DIRECTIONAL CELSIUS FLTR CATHETERS (D-1359-XX-S)
Applicant
Biosense Webster, Inc.
PMA number
P950005
Supplement
S041
Product code
LPB
Decision date
2013-06-06
Classification
Cardiac Ablation Percutaneous Catheter
Generic name
Cardiac ablation percutaneous catheter
Approval order statement
APPROVAL FOR MODIFIED CATHETER DESIGNS TO THE 7FR CELSIUS AND 7FR EZ STEER DS CATHETERS TO CREATE THE 7.5FR 8 MM CELSIUS FLTR CATHETERS. THE CHANGES BEING APPROVED ARE THE INCREASED DISTAL PROFILE, LARGER CURVE PROFILE, INCREASED RANGE OF DEFLECTION, AND INCREASED DEFLECTABLE TIP DIAMETER. THE DEVICES, AS MODIFIED, WILL BE MARKETED UNDER THE TRADE NAMES BI-DIRECTIONAL CELSIUS FLTR CATHETERS (D-1359-XX-S) AND UNI-DIRECTIONAL CELSIUS FLTR CATHETERS (D-1355-XX-S) AND ARE INDICATED FOR CATHETER-BASED CARDIAC ELECTROPHYSIOLOGICAL MAPPING (STIMULATION AND RECORDING), AND WHEN USED WITH THE STOCKERT 70 (WITH SOFTWARE VERSION 001/033 OR HIGHER) FOR THE TREATMENT OF TYPE I ATRIAL FLUTTER IN PATIENTS AGE 18 YEARS OF AGE OR OLDER.

Current openFDA PMA Record#

Device
BI-DIRECTIONAL CELSIUS FLTR CATHETERS (D-1359-XX-S)
Applicant
Biosense Webster, Inc.
PMA number
P950005
Supplement
S041
Product code
LPB
Generic name
Cardiac ablation percutaneous catheter
Decision date
2013-06-06
Decision code
APPR
Date received
2012-11-07
Supplement type
Normal 180 Day Track
Supplement reason
Change Design/Components/Specifications/Material
Approval order statement
APPROVAL FOR MODIFIED CATHETER DESIGNS TO THE 7FR CELSIUS AND 7FR EZ STEER DS CATHETERS TO CREATE THE 7.5FR 8 MM CELSIUS FLTR CATHETERS. THE CHANGES BEING APPROVED ARE THE INCREASED DISTAL PROFILE, LARGER CURVE PROFILE, INCREASED RANGE OF DEFLECTION, AND INCREASED DEFLECTABLE TIP DIAMETER. THE DEVICES, AS MODIFIED, WILL BE MARKETED UNDER THE TRADE NAMES BI-DIRECTIONAL CELSIUS FLTR CATHETERS (D-1359-XX-S) AND UNI-DIRECTIONAL CELSIUS FLTR CATHETERS (D-1355-XX-S) AND ARE INDICATED FOR CATHETER-BASED CARDIAC ELECTROPHYSIOLOGICAL MAPPING (STIMULATION AND RECORDING), AND WHEN USED WITH THE STOCKERT 70 (WITH SOFTWARE VERSION 001/033 OR HIGHER) FOR THE TREATMENT OF TYPE I ATRIAL FLUTTER IN PATIENTS AGE 18 YEARS OF AGE OR OLDER.