PMA P960040S440

Device
Autogen, Dynagen, Inogen, and Origen NG3 ICDs and Resonate, Momentum, Vigilant, and Perciva NG4 ICDs
Applicant
Boston Scientific
PMA number
P960040
Supplement
S440
Product code
LWS
Decision date
2019-09-18
Classification
Implantable Cardioverter Defibrillator (non-crt)
Generic name
Implantable cardioverter defibrillator (non-CRT)
Approval order statement
Approval for hardware modifications to the IS-1 lead bore cavity and spring coil in the IS-1 pulse generator (PG) header for all pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps) within the Accolade family and a subset of implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-D) within the NG3 and NG4 families.

Current openFDA PMA Record#

Device
Autogen, Dynagen, Inogen, and Origen NG3 ICDs and Resonate, Momentum, Vigilant, and Perciva NG4 ICDs
Applicant
Boston Scientific
PMA number
P960040
Supplement
S440
Product code
LWS
Generic name
Implantable cardioverter defibrillator (non-CRT)
Decision date
2019-09-18
Decision code
APPR
Date received
2019-06-20
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Approval order statement
Approval for hardware modifications to the IS-1 lead bore cavity and spring coil in the IS-1 pulse generator (PG) header for all pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps) within the Accolade family and a subset of implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-D) within the NG3 and NG4 families.