This medical device record is a pma supplement. A supplement may have changed the device description/function or indication from that approved in the original pma. Be sure to look at the original pma record for more information.
The 30-day notice requested that fda review a change in the manufacturing process. A vertrod thermal impulse heat sealer will be used in place of an audion future jaw clamp pouch sealer to seal the tyvek pouch containing the aluminum tube of adcon(r)-l adhesion control barrier gel. This change does not affect the currently approved tyvek seal integrity testing method or specification.
Device | ADCON-L ANTI-ADHESION CONTROL IN A BARRIER GEL |
Classification Name | Inhibitor, Peridural Fibrosis (adhesion Barrier) |
Generic Name | Inhibitor, Peridural Fibrosis (adhesion Barrier) |
Applicant | WRIGHTMEDICALTECHNOLOGYINC |
Date Received | 1998-08-28 |
Decision Date | 1998-09-11 |
PMA | P960057 |
Supplement | S001 |
Product Code | MLQ |
Advisory Committee | Physical Medicine |
Supplement Type | 30-day Notice |
Supplement Reason | Process Change - Manufacturer/sterilizer/packager/supplier |
Expedited Review | No |
Combination Product | No |
Applicant Address | WRIGHTMEDICALTECHNOLOGYINC 5677 Airline Rd. arlinton, TN 38002 |
Supplement Number | Date | Supplement Type |
---|---|---|
P960057 | Original Filing | |
S011 | 2000-06-26 | Special (immediate Track) |
S010 | ||
S009 | ||
S008 | 1999-05-13 | Special (immediate Track) |
S007 | ||
S006 | 1999-03-12 | Real-time Process |
S005 | 1999-02-19 | 135 Review Track For 30-day Notice |
S004 | 1999-02-11 | Normal 180 Day Track |
S003 | 1998-12-21 | 30-day Notice |
S002 | 1998-11-03 | Real-time Process |
S001 | 1998-08-28 | 30-day Notice |