- Device
- LADAR VISION TREATMENTS
- Applicant
- Alcon Laboratories
- PMA number
- P970043
- Supplement
- S004
- Product code
- LZS
- Generic name
- Excimer laser system
- Decision date
- 1999-09-24
- Decision code
- APPR
- Date received
- 1999-08-19
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Approval order statement
- Approval for adding software computer screens to provide precautions when refractive parameters were entered that were outside the approved indications for use. These indications for use remain: PRK for the reduction or elimination of -1.00 to -10.00 D of spherical myopia with or without -1.00 to -4.00 D of astigmatism at the spectacle plane, the combination of which must result in an attempted correction between -0.50 and -10.00 D SE at the spectacle plane where sphere or cylinder magnitude is at least 1.00 D.