- Device
- OVATIO DR MODEL 6550 AND OVATIO VR MODEL 6250 ICDS, ELAVIEW 1.32 UG2 PROGRAMMING SOFTWARE AND CPR3 PROGRAMMING HEAD
- Applicant
- MicroPort CRM USA, Inc.
- PMA number
- P980049
- Supplement
- S020
- Product code
- MRM
- Generic name
- DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER
- Decision date
- 2006-04-28
- Decision code
- APPR
- Date received
- 2005-04-29
- Supplement type
- Normal 180 Day Track
- Supplement reason
- Change Design/Components/Specifications/Material
- Approval order statement
- APPROVAL FOR AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT COMBINES FEATURES OF THE ALTO 2 DR 624, ALTO 2 VR 625 ICDS, AND SYMPHONY 2550 PACEMAKER, INCLUDING REDUCED SIZE, MODIFIED SHOCK CIRCUITRY AND OTHER CHANGES. THE DEVICE WILL BE MARKETED UNDER THE TRADE NAME OVATIO DR MODEL 6550 AND OVATIO VR MODEL 6250 ICDS, ELAVIEW 1.32 UG2 PROGRAMMING SOFTWARE AND CPR3 PROGRAMMING HEAD WITH HSO 2.76 SOFTWARE AND IS INDICATED FOR USE IN PATIENTS WHO ARE AT HIGH RISK OF SUDDEN CARDIAC DEATH DUE TO VENTRICULAR TACHYARRHYTHMIAS AND WHO HAVE EXPERIENCED ONE OF THE FOLLOWING SITUATIONS: 1) SURVIVAL OF AT LEAST ONE EPISODE OF CARDIAC ARREST (MANIFESTED BY THE LOSS OF CONSCIOUSNESS) DUE TO VENTRICULAR TACHYARRHYTHMIA, 2) RECURRENT, POORLY TOLERATED SUSTAINED VENTRICULAR TACHYCARDIA (VT).NOTE: THE CLINICAL OUTCOME FOR HEMODYNAMICALLY STABLE VT PATIENTS IS NOT FULLY KNOWN. SAFETY AND EFFECTIVENESS STUDIES HAVE NOT BEEN CONDUCTED.