Recall Z-0073-2027

Recall number
Z-0073-2027
Event number
99704
Firm
Steris Corporation
Firm FEI
1527821
Product code
FRC
Status
Open, Classified
Initiated
2026-08-31
Posted
2026-10-02
Root cause
Under Investigation by firm
510(k) numbers
K083665, K191021, K113221, K963841, K240717, K223715, K895704, K111391, K173584, K891784, K926364, K190297, K844479, K172474, K971432, K140392, K031012, K771080, K771132, K855120, K140620, K101910, K200766, K032914, K930682, K250044, K971430, K073244, K122362, K812272

Product#

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

Reason for Recall#

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction