Recall Z-0084-2027

Recall number
Z-0084-2027
Event number
99800
Firm
Beckman Coulter Inc.
Firm FEI
2050012
Product code
JJE
Status
Open, Classified
Initiated
2026-09-02
Posted
2026-10-02
Root cause
Software design
510(k) numbers
K220977, K950915, K953239, K831518, K970227, K130915, K803267, K862818, K863840, K901346, K922220, K881498, K813102, K840883, K841916, K855196, K921661, K922072, K834521, K870833, K942443, K782014, K010500, K885226, K944028, K860748, K912844, K110035, K852262, K881495

Product#

DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520

Reason for Recall#

Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.