Recall Z-0089-2027

Recall number
Z-0089-2027
Event number
99549
Firm
Nihon Kohden America LLC
Firm FEI
2080783
Product code
MHX
Status
Open, Classified
Initiated
2026-09-08
Posted
2026-10-06
Root cause
Software design
510(k) numbers
K102376, K102040, K033711, K093268, K033600, K000854, K113653, K061185, K000517, K173768, K100040, K192514, K080495, K252250, K040395, K942700, K051400, K132077, K040955, K990068, K040357, K073462, K131414, K131829, K001722, K250349, K082045, K950264, K260579, K062095

Product#

Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No

Reason for Recall#

Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.