Recall Z-0141-2027

Recall number
Z-0141-2027
Event number
99774
Firm
Cadwell Industries Inc
Firm FEI
3020018
Product code
GWF
Status
Open, Classified
Initiated
2026-08-18
Posted
2026-10-09
Root cause
Under Investigation by firm
510(k) numbers
K162199, K883340, K992911, K912685, K863175, K190703, K884870, K010087, K903731, K860903, K112052, K843253, K162873, K950208, K991899, K071821, K971740, K811848, K870567, K900482, K810482, K992313, K001101, K082614, K212166, K860240, K822744, K831896, K912775, K010090

Product#

Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system

Reason for Recall#

Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.