Recall Z-0208-2027

Recall number
Z-0208-2027
Event number
99778
Firm
Abiomed, Inc.
Firm FEI
1220648
Product code
OZD
Status
Open, Classified
Initiated
2026-09-18
Posted
2026-10-28
Root cause
Under Investigation by firm

Product#

Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump.

Reason for Recall#

Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.