Recall Z-0315-2027

Recall number
Z-0315-2027
Event number
99900
Firm
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Firm FEI
3007420694
Product code
FSA
Status
Open, Classified
Initiated
2026-09-25
Posted
2026-10-29
Root cause
Under Investigation by firm
510(k) numbers
K930625, K841398, K792601, K942807, K811899, K780684, K823206, K902141, K844976, K860837, K840997, K841395, K851844

Product#

Carevo, Height-adjustable shower trolley, Model Number BAC1111-01

Reason for Recall#

The side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient handling. If the side support is not fully locked and an additional force is applied by the caregiver on the handle (e.g., during turning or repositioning), the panel may open unintentionally, which can lead to a resident/patient fall and serious injury, such as head injury, fractures, or other trauma resulting from a patient fall.