Recall Z-0317-2027
- Recall number
- Z-0317-2027
- Event number
- 99863
- Firm
- Accriva Diagnostics, Inc.
- Firm FEI
- 2250033
- Product code
- JPA
- Status
- Open, Classified
- Initiated
- 2026-09-14
- Posted
- 2026-10-30
- Root cause
- Process control
- 510(k) numbers
- K940401, K161312, K761172, K060162, K000973, K221359, K854938, K914188, K994349, K761171, K193041, K883689, K050513, K951891, K002177, K853986, K162420, K930849, K923761, K000053, K931149, K761091, K212183, K894359, K885076, K930068, K924509, K883484, K183160, K093167
Product#
Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103
Reason for Recall#
Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.