Recall Z-0317-2027

Recall number
Z-0317-2027
Event number
99863
Firm
Accriva Diagnostics, Inc.
Firm FEI
2250033
Product code
JPA
Status
Open, Classified
Initiated
2026-09-14
Posted
2026-10-30
Root cause
Process control
510(k) numbers
K940401, K161312, K761172, K060162, K000973, K221359, K854938, K914188, K994349, K761171, K193041, K883689, K050513, K951891, K002177, K853986, K162420, K930849, K923761, K000053, K931149, K761091, K212183, K894359, K885076, K930068, K924509, K883484, K183160, K093167

Product#

Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103

Reason for Recall#

Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.