Recall Z-0319-2027
- Recall number
- Z-0319-2027
- Event number
- 99941
- Firm
- NATUS MEDICAL INCORPORATED
- Firm FEI
- 3002809369
- Product code
- MNR
- Status
- Open, Classified
- Initiated
- 2026-09-29
- Posted
- 2026-10-30
- Root cause
- Component change control
- 510(k) numbers
- K122516, K222579, K041904, K222950, K973920, K200654, K042745, K183625, K203839, K081485, K203343, K072014, K201495, K183559, K024322, K082853, K254042, K982293, K244027, K231355, K021138, K041724, K210480, K191031, K010156, K180001, K102567, K243220, K990199, K242290
Product#
Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011
Reason for Recall#
The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.