Recall Z-0319-2027

Recall number
Z-0319-2027
Event number
99941
Firm
NATUS MEDICAL INCORPORATED
Firm FEI
3002809369
Product code
MNR
Status
Open, Classified
Initiated
2026-09-29
Posted
2026-10-30
Root cause
Component change control
510(k) numbers
K122516, K222579, K041904, K222950, K973920, K200654, K042745, K183625, K203839, K081485, K203343, K072014, K201495, K183559, K024322, K082853, K254042, K982293, K244027, K231355, K021138, K041724, K210480, K191031, K010156, K180001, K102567, K243220, K990199, K242290

Product#

Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011

Reason for Recall#

The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.