Recall Z-0322-2027
- Recall number
- Z-0322-2027
- Event number
- 99957
- Firm
- Illumina, Inc.
- Firm FEI
- 3003218906
- Product code
- PFF
- Status
- Open, Classified
- Initiated
- 2024-05-08
- Posted
- 2026-10-30
- Root cause
- Labeling design
- 510(k) numbers
- K170299
Product#
NovaSeq 6000Dx Instrument REF: 20068232, FOR IN VITRO DIAGNOSTIC USE
Reason for Recall#
(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission.