Recall Z-0333-2027
- Recall number
- Z-0333-2027
- Event number
- 99826
- Firm
- Baxter Healthcare Corporation
- Firm FEI
- 1416980
- Product code
- GDC
- Status
- Open, Classified
- Initiated
- 2026-09-14
- Posted
- 2026-11-03
- Root cause
- Equipment maintenance
- 510(k) numbers
- K971390, K883613, K851830, K851021, K850168, K821665, K964273, K843134, K821749, K940616, K915393, K914758, K853480, K934744, K970807, K851829, K853720, K853535
Product#
Mobile column TruSystem 7500 U, Product Code 1730720
Reason for Recall#
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.