Recall Z-0348-2027
- Recall number
- Z-0348-2027
- Event number
- 100047
- Firm
- restor3d Inc.
- Firm FEI
- 3014833750
- Product code
- PHX
- Status
- Open, Classified
- Initiated
- 2026-10-01
- Posted
- 2026-11-03
- Root cause
- Device Design
- 510(k) numbers
- K243643, K121183, K180632, K261016, K202098, K234105, K050316, K212415, K053274, K254003, K210533, K150568, K113254, K152754, K063569, K182536, K193394, K251184, K130550, K214001, K111061, K232583, K133378, K260556, K100741, K052906, K182039, K093803, K161120, K211074
Product#
Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner
Reason for Recall#
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.