Recall Z-0348-2027

Recall number
Z-0348-2027
Event number
100047
Firm
restor3d Inc.
Firm FEI
3014833750
Product code
PHX
Status
Open, Classified
Initiated
2026-10-01
Posted
2026-11-03
Root cause
Device Design
510(k) numbers
K243643, K121183, K180632, K261016, K202098, K234105, K050316, K212415, K053274, K254003, K210533, K150568, K113254, K152754, K063569, K182536, K193394, K251184, K130550, K214001, K111061, K232583, K133378, K260556, K100741, K052906, K182039, K093803, K161120, K211074

Product#

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

Reason for Recall#

(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.