Recall Z-0352-2027

Recall number
Z-0352-2027
Event number
100117
Firm
Baxter Healthcare Corporation
Firm FEI
1417572
Product code
GDC
Status
Open, Classified
Initiated
2026-09-29
Posted
2026-11-03
Root cause
Nonconforming Material/Component
510(k) numbers
K971390, K883613, K851830, K851021, K850168, K821665, K964273, K843134, K821749, K940616, K915393, K914758, K853480, K934744, K970807, K851829, K853720, K853535

Product#

Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788

Reason for Recall#

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.