Recall Z-2735-2026

Recall number
Z-2735-2026
Event number
99304
Firm
ResMed Ltd.
Firm FEI
3004604967
Product code
CBK
Status
Open, Classified
Initiated
2026-06-25
Posted
2026-07-28
Root cause
Component design/selection
510(k) numbers
K152068, K172875, K822456, K844350, K912619, K840119, K252663, K781160, K960861, K871792, K241520, K911336, K113051, K922705, K010317, K073069, K980642, K102775, K900314, K821930, K110083, K111610, K895541, K823861, K923444, K183364, K100753, K063806, K861707, K864105

Product#

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

Reason for Recall#

Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.