Recall Z-2771-2026

Recall number
Z-2771-2026
Event number
99236
Firm
Boston Scientific Corporation
Firm FEI
3002095335
Product code
LLZ
Status
Open, Classified
Initiated
2026-06-19
Posted
2026-07-22
Root cause
Device Design
510(k) numbers
K252547, K260479, K073169, K050619, K103565, K042212, K070683, K053183, K131873, K161625, K022970, K151941, K202256, K201202, K163013, K173101, K093982, K000016, K210187, K112354, K992654, K120525, K161109, K212421, K960043, K032096, K081556, K202292, K042647, K242647

Product#

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

Reason for Recall#

Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.