Recall Z-2775-2026

Recall number
Z-2775-2026
Event number
99406
Firm
KENT IMAGING, INC.
Firm FEI
3009771004
Product code
QJF
Status
Open, Classified
Initiated
2026-06-30
Posted
2026-07-22
Root cause
Software design
510(k) numbers
K242669, K210882, K213840, K230734, K251678, K191371, K211901

Product#

SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.

Reason for Recall#

The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.