Recall Z-2805-2026

Recall number
Z-2805-2026
Event number
99267
Firm
Permobil Inc
Firm FEI
1221084
Product code
ITI
Status
Open, Classified
Initiated
2025-06-19
Posted
2026-07-24
Root cause
Process design
510(k) numbers
K123290, K142770, K863233, K001777, K062888, K151717, K904862, K933856, K013516, K123769, K900037, K024175, K011687, K050010, K122357, K160139, K193599, K070433, K010075, K852810, K911152, K070655, K032512, K905292, K010380, K232193, K972230, K041208, K800209, K010941

Product#

M5 Corpus Power Wheelchair REF,120032

Reason for Recall#

This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.