Recall Z-3085-2026

Recall number
Z-3085-2026
Event number
99556
Firm
Fresenius Medical Care Renal Therapies Group, LLC
Firm FEI
3001451489
Product code
KOC
Status
Open, Classified
Initiated
2026-08-10
Posted
2026-09-02
Root cause
Component change control
510(k) numbers
K243505, K760626, K760935, K760640, K823069, K925522, K950282, K823707, K761051, K770048, K230514, K811457, K832666, K770383, K761082, K831175, K172238, K913361, K924436, K830231, K831482, K941803, K844185, K760042, K831338, K790591, K831433, K833252, K834596, K823791

Product#

5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3

Reason for Recall#

Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.