melphalan hydrochloride injection, powder, lyophilized, for solution
- Manufacturer
- Delcath Systems, Inc.
- Effective date
- 2025-09-04
- Label type
- Risk Evaluation & Mitigation Strategies
- Version
- 8
- Source
- full-release
- Hydrated at
- 2026-06-04 22:50:43
Label at a glance#
Contacts and useful links#
Label contents#
Full prescribing information#
I. Administrative Information
Risk: severe peri-procedural complications including hemorrhage, hepatocellular injury, and thromboembolic events
Application Number: NDA 201848
Application Holder: Delcath Systems, Inc.
Initial REMS Approval: 08/2023
Most Recent REMS Update: 9/2025
II. REMS Goal
The goal of the HEPZATO KIT REMS is to mitigate the risks of severe peri-procedural complications including hemorrhage, hepatocellular injury, and thromboembolic events associated with HEPZATO KIT.
Objective: Minimize severe peri-procedural complications
III. REMS Requirements
REMS PARTICIPANT REQUIREMENTS
Delcath Systems, Inc. must ensure that healthcare settings comply with the following requirements:
| 1. Healthcare settings that dispense HEPZATO KIT must: | |
| To become certified to dispense | 1. Have percutaneous hepatic perfusion procedure team(s) that must include healthcare providers with expertise in interventional radiology, anesthesiology, and perfusion as described in the Instructions for Use. 2. Have the following on-site: interventional radiology suite or operating room with fluoroscopy and with resuscitation personnel, equipment, and medications. 3. Designate an Authorized Representative to carry out the certification process and oversee implementation and compliance with the REMS requirements on behalf of the healthcare setting. 4. Have the Authorized Representative review the product's Prescribing Information, Instructions for Use, Program Overview, and Didactic Modules. 5. Have the Authorized Representative enroll in the REMS by completing and submitting the Healthcare Setting Enrollment Form to the REMS. 6. Have the percutaneous hepatic perfusion procedure team members that perform procedures with HEPZATO KIT review the product's Prescribing Information and Instructions for Use. 7. Have the percutaneous hepatic perfusion procedure team members that perform procedures with HEPZATO KIT take the Program Overview, Didactic Modules, and Preceptorship training provided by Delcath Systems, Inc. 8. Have the percutaneous hepatic perfusion procedure team members that perform procedures with HEPZATO KIT successfully complete the Proctorship training provided by Delcath Systems, Inc., and successfully complete and submit the Criteria for Procedural Competency Checklist to the REMS. 9. Establish processes and procedures to ensure relevant new staff who will perform percutaneous hepatic perfusion procedures with HEPZATO KIT are trained. |
| Before administering | 10. Obtain authorization to dispense each HEPZATO KIT by contacting the REMS to verify the percutaneous hepatic perfusion procedure team is qualified using the Procedure Team Qualification Status Form. |
| During and after administering, for at least 72 hours | 11. Assess the patient for severe peri-procedural complications associated with HEPZATO KIT. Document and submit severe peri-procedural complications using the Severe Peri-Procedure-related Complications Adverse Events Documentation Form to the REMS. |
| To maintain certification to dispense, annually | 12. Have percutaneous hepatic perfusion procedure team(s) that must include healthcare providers with expertise in interventional radiology, anesthesiology, and perfusion as described in the Instructions For Use. 13. Have percutaneous hepatic perfusion procedure team members who have each performed one procedure in the first six months following completion of training, a second procedure in the next six months, and at least two procedures annually thereafter. 14. Have the following on-site: interventional radiology suite or operating room with fluoroscopy and with resuscitation personnel, equipment, and medications. 15. If there is a change in Authorized Representative, have a new Authorized Representative enroll in the REMS by completing the Healthcare Setting Enrollment Form. |
| At all times | 16. Maintain records of each percutaneous hepatic perfusion procedure team member's training. 17. Maintain records of the percutaneous hepatic perfusion procedures performed with HEPZATO KIT and the associated percutaneous hepatic perfusion procedure team members' participation. 18. Maintain records that processes and procedures are in place and are being followed. 19. Comply with audits carried out by Delcath Systems, Inc., or a third party acting on behalf of Delcath Systems, Inc., to ensure that all training, processes, and procedures are in place and are being followed. |
REMS APPLICANT REQUIREMENTS
Delcath Systems, Inc. must provide training to healthcare settings that dispense HEPZATO KIT.
The training includes the following educational materials: Program Overview, Didactic Modules. The training must be made available in hardcopy format and must be presented in-person or via live webcast.
The training also includes a Preceptorship (a live, in-person, observation experience) and a Proctorship (performing a live, in-person, percutaneous hepatic perfusion procedure with HEPZATO KIT observed and evaluated by a Delcath Systems, Inc. representative) documented with a Criteria for Procedural Competency Checklist.
To support REMS operations, Delcath Systems, Inc. must:
- Ensure the HEPZATO KIT is only distributed to certified healthcare settings.
-
Authorize dispensing for each procedure based on receipt of the Procedure Team Qualification Status Form to verify the percutaneous hepatic perfusion procedure team is qualified to perform procedures with HEPZATO KIT. The percutaneous hepatic perfusion procedure team is qualified if it includes healthcare providers with expertise in interventional radiology, anesthesiology, and perfusion as described in the Instructions For Use who have each successfully completed the training and performed one procedure in the first six months following completion of training, a second procedure in the next six months, and at least two procedures annually thereafter.
If the Procedure Team Qualification Status Form is not received and these criteria are not met, the healthcare setting is not authorized to dispense and administer HEPZATO KIT. - Establish and maintain a process to ensure that healthcare settings that want to become certified to dispense HEPZATO KIT and to maintain certification have a percutaneous hepatic perfusion procedure team with at least one healthcare provider with expertise in interventional radiology, anesthesiology, and perfusion, as described in the Instructions For Use.
-
Establish and maintain a REMS website,
www.HEPZATOKITREMS.com. The REMS Website must include the option to complete all healthcare setting forms online and the option to print the Prescribing Information including Instructions For Use, and REMS materials. All product websites for consumers and healthcare providers must include prominent REMS-specific links to the REMS Website. The REMS website must not link back to the promotional product website(s).
- Make the REMS website fully operational and all REMS materials available through the REMS Website or REMS Coordinating Center at the time HEPZATO KIT first becomes commercially available.
- Establish and maintain a REMS Coordinating Center for the REMS participants 1-833-632-0457.
- Establish and maintain a validated, secure database of all REMS participants who are enrolled and/or certified in the REMS.
- Ensure healthcare settings are able to enroll in the REMS via e-mail and online.
- Ensure healthcare settings are able to obtain authorization to dispense and administer HEPZATO KIT by e-mail.
- Notify healthcare settings within 7 calendar days after they become certified in the REMS.
- Provide public access to a database of certified healthcare settings.
- Report severe peri-procedural complications including hemorrhage, hepatocellular injury, and thromboembolic events associated with HEPZATO KIT as soon as possible to the FDA but no later than 15 calendar days from the initial receipt of the information by the applicant. This requirement does not affect the applicant's other reporting and follow-up requirements under applicable FDA regulations.
To ensure REMS participants' compliance with the REMS, Delcath Systems, Inc. must:
- Verify annually that the certified healthcare setting has at least one qualified percutaneous hepatic perfusion procedure team(s) that includes at least one healthcare provider with expertise in interventional radiology, anesthesiology, and perfusion as described in the Instructions For Use and have performed one procedure in the first six months following completion of training, a second procedure in the next six months, and at least two procedures annually thereafter. If these qualifications are not met, the healthcare setting must have the healthcare provider(s) with the necessary expertise as described in the Instructions For Use complete the Didactic Modules and/or Proctorship training to re-certify.
- Verify annually that the designated Authorized Representative's name and contact information correspond to those of the current designated Authorized Representative for the healthcare setting. If different, the healthcare setting must re-certify with a new Authorized Representative.
- Establish and maintain a process to retrain percutaneous hepatic perfusion procedure team members who have not performed one procedure in the first six months following completion of the initial training, a second procedure in the next six months, and at least two procedures annually thereafter. The re-training must include the Didactic Modules. The Proctorship training must also be completed by a percutaneous hepatic perfusion procedure team member who has not performed a procedure with HEPZATO KIT within the preceding twelve months.
- Maintain adequate records to demonstrate that REMS requirements have been met, including, but not limited to records of: HEPZATO KIT distribution and dispensing; certification of healthcare settings, and audits of REMS participants. These records must be readily available for FDA inspections.
- Establish and maintain a plan for addressing noncompliance with REMS requirements.
- Monitor healthcare settings on an ongoing basis to ensure the requirements of the REMS are being met. Take corrective action if non-compliance is identified, including de-certification.
- Audit annually: 1.) healthcare settings no later than 180 calendar days after the healthcare setting certification date and annually thereafter and 2.) the REMS Coordinating Center to ensure that all REMS processes and procedures are in place, functioning, and support the REMS requirements.
- Take reasonable steps to improve operations of and compliance with the requirements in the HEPZATO KIT REMS based on monitoring and evaluation of the HEPZATO KIT REMS.
IV. REMS Assessment Timetable
Delcath Systems, Inc. must submit REMS Assessments at 6 months, 12 months, and annually thereafter from the date of the initial approval of the REMS (08/14/2023). To facilitate inclusion of as much information as possible while allowing reasonable time to prepare the submission, the reporting interval covered by each assessment should conclude no earlier than 60 calendar days before the submission date for that assessment. Delcath Systems, Inc. must submit each assessment so that it will be received by the FDA on or before the due date.
V. REMS Materials
The following materials are part of the HEPZATO KIT REMS:
Enrollment Forms
Healthcare Setting:
- Healthcare Setting Enrollment Form
Training and Educational Materials
Healthcare Setting:
- Program Overview
- Didactic Modules
- Criteria for Procedural Competency Checklist
Patient Care Form
- Severe Peri-Procedure-related Complications Adverse Events Documentation Form
Other Materials
- Procedure Team Qualification Status Form
- REMS Website
VI. Statutory Elements
This REMS is required under section 505-1 of the Federal Food, Drug, and Cosmetic Act (FD&C Act) (21 U.S.C. 355-1) and consists of the following elements:
-
Elements to Assure Safe Use:
- Health care settings that dispense HEPZATO KIT are specially certified under 505-1(f)(3)(B)
- HEPZATO KIT is dispensed to patients only in certain health care settings under 505-1(f)(3)(C)
- HEPZATO KIT is dispensed to patients with evidence or other documentation of safe-use conditions under 505-1(f)(3)(D)
- Each patient using HEPZATO KIT is subject to certain monitoring under 505-1(f)(3)(E)
- Implementation System
- Timetable for Submission of Assessments
DailyMed Product Concepts#
DailyMed REMS / Drug Safety Documents#
Labels related to this REMS document
| DailyMed label | Relation | Version | Effective | Manufacturer | Source |
|---|---|---|---|---|---|
| These highlights do not include all the information needed to use HEPZATO safely and effectively. See full prescribing information for HEPZATO KIT and HEPZATO KIT Hepatic Delivery System Instructions for Use. HEPZATO (melphalan) for injection is a component of the HEPZATO KIT Hepatic Delivery System (HDS) for intra-arterial use. Initial U.S. Approval: 1964 | SPL related document | 2026-06-09 | Delcath Systems, Inc. | 20251107_01291f98-96ef-4336-95f8-825dd19690ce.zip!18a9ad85-1d15-4d41-aeab-1cab61a20f14.xml |
Products#
Every source-derived product name is available through these pages.
- Hepzato Kit
- melphalan hydrochloride injection, powder, lyophilized, for solution
- Hepzato Drug
- melphalan hydrochloride
- melphalan
- Hepzato Diluent
- sodium citrate, propylene glycol, alcohol, and water
- Hepzato Secondary Diluent
- sodium chloride and water
- Sodium Chloride
- Sodium Cation
- Delcath Systems, Inc.
NDC Codes#
Ingredients#
Every source-derived ingredient row is available through these pages.
| Name | UNII | Kind |
|---|---|---|
| Melphalan Hydrochloride | 1VXP4V453T | ACTIM |
| Sodium Chloride | 451W47IQ8X | ACTIB |
Source Document#
openFDA label cross-check#
OpenFDA label data provides additional search and identifier links. DailyMed’s Structured Product Label is the canonical label on FDA.report. Matching records are deduplicated before they are shown below.
| Brand | Generic | Manufacturer | SPL set ID | Effective date | Available safety fields | Join |
|---|---|---|---|---|---|---|
| Melphalan hydrochloride | MELPHALAN HYDROCHLORIDE | Heritage Pharmaceuticals Inc. d/b/a Avet Pharmaceuticals Inc. | 2c18beeb-7bd4-4591-94cb-9d7156c1b8e0 | 2026-07-07 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Hepzato Kit | MELPHALAN HYDROCHLORIDE INJECTION, POWDER, LYOPHILIZED, FOR SOLUTION | Delcath Systems, Inc. | 4f83c8f7-4cc0-4219-88d5-7cfddce91198 | 2026-06-09 | Boxed warning, Warnings, Adverse reactions | Hepzato Kit |
| Melphalan | MELPHALAN HYDROCHLORIDE | Fresenius Kabi USA, LLC | e9c2efee-173d-4795-96d8-7aae5b3a3073 | 2026-01-14 | Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan Hydrochloride | MELPHALAN HYDROCHLORIDE | Arthur Group LLC | 2254450e-4803-b947-e063-6394a90a4624 | 2026-01-08 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan hydrochloride | MELPHALAN HYDROCHLORIDE | BPI Labs LLC | dab93975-0587-4f12-848c-9130bb02d036 | 2025-12-05 | Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan Hydrochloride | MELPHALAN HYDROCHLORIDE | Qilu Pharmaceutical Co., Ltd. | 91948ff9-6070-485b-b4b7-3cb477c1b24c | 2024-07-04 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| MELPHALAN HYDROCHLORIDE | MELPHALAN HYDROCHLORIDE | Camber Pharmaceuticals, Inc. | 24006cbf-4e91-4bb2-9627-25b4ebc9470a | 2024-06-21 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan Hydrochloride | MELPHALAN HYDROCHLORIDE | Sagent Pharmaceuticals | 35f3d9fe-3c2f-4362-9d40-cdab014cac38 | 2024-03-25 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan Hydrochloride | MELPHALAN HYDROCHLORIDE | Dr.Reddy's Laboratories Inc | bd5e4db5-8787-4428-2cb6-ef6f8a5e08af | 2024-02-27 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan Hydrochloride | MELPHALAN HYDROCHLORIDE | Avenacy Inc. | 8edd78a9-d762-4578-a610-28d6595f4bf3 | 2023-12-18 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan hydrochloride | MELPHALAN HYDROCHLORIDE | Meitheal Pharmaceuticals Inc. | d99b3865-e0f0-482a-9ec6-da642e5104ea | 2021-06-07 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan Hydrochloride | MELPHALAN HYDROCHLORIDE | Mylan Institutional LLC | a8bf1877-fd06-4ef6-b82d-fe0ff1956bde | 2020-09-16 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan hydrochloride | MELPHALAN HYDROCHLORIDE | Fosun Pharma USA Inc | 54712714-744d-4deb-83b2-694da49a9272 | 2019-07-28 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan hydrochloride | MELPHALAN HYDROCHLORIDE | Gland Pharma Limited | ea7fbf4e-4980-4bb4-bac9-99eb6f633ba0 | 2019-06-07 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
| Melphalan Hydrochloride | MELPHALAN HYDROCHLORIDE | Dr.Reddy's Laboratories Inc | d4705154-3050-1a5c-950f-e24b51487401 | 2018-04-10 | Boxed warning, Warnings, Adverse reactions | MELPHALAN HYDROCHLORIDE |
Reported adverse events (FAERS/openFDA)#
Adverse event summaries are temporarily unavailable. Other product information remains available.