Leukotrap WB System

Manufacturer
Haemonetics Corporation | Haemonetics Manufacturing Inc
Effective date
2020-01-02
Label type
HUMAN PRESCRIPTION DRUG LABEL
Version
4
Source
full-release
Hydrated at
2026-05-31 20:31:14

Label at a glance#

ProductLeukotrap WB System
Active ingredientDEXTROSE MONOHYDRATE, SODIUM CITRATE, CITRIC ACID MONOHYDRATE, SODIUM PHOSPHATE, MONOBASIC, MONOHYDRATE, ADENINE
Label structure13 sections

Indications and uses

Indicated for collection of blood and preparation of red blood cells and plasma with pre-storage leukocyte reduction. Platelet concentrates are not intended to be made with this product.

Label contents#

Full prescribing information#

DESCRIPTION

DESCRIPTION SECTION

CPDA-1 Blood Collection Set, Double Bag with WBF2 Filter and Sampling System

Instruction for Use for Systems Containing a Y Sampling Site (YSS) or Sample Diversion Pouch (with or without a pre-attached Samp Lok Vacuum Tube Holder). Refer to unit foil package label for specific product description being used.

Sterile, nonpyrogenic fluid path. Sterilized by steam.

Rx only.

This product is free of natural rubber latex.

INDICATIONS AND USAGE

INDICATIONS & USAGE SECTION

Indicated for collection of blood and preparation of red blood cells and plasma with pre-storage leukocyte reduction. Platelet concentrates are not intended to be made with this product.

WARNINGS

WARNINGS SECTION

Failure to achieve and maintain a closed system during processing would result in a product that must be transfused within 24 hours.

GENERALPRECAUTIONS

PRECAUTIONS SECTION

Use Aseptic Technique.

Use only if solution is clear.

Store CPDA-1 preserved red blood cells at 1—6 °C for up to 35 days and use as indicated.

* During processing, always observe the following precautions:

1. Sealing should be done in a manner that avoids fluid splatter.

2. Always dispose of blood-contaminated products in a manner consistent with established BIOHAZARD safety procedures.

ADVERSE REACTIONS

ADVERSE REACTIONS SECTION

There are no known potential adverse reactions associated specifically with the Leukotrap WB System with CPDA-1.

I. BLOOD COLLECTION INSTRUCTIONS FOR SYSTEMS CONTAINING A Y SAMPLING SITE (YSS) ONLY

SPL UNCLASSIFIED SECTION

1. Load blood agitation device or suspend blood bag on donor scale and adjust donor scale to desired collection gross weight as per manufacturer's instructions.

2. Clamp donor tubing between the Donor Care® Needle Guard (DCNG) and Y Sampling Site.

3. Secure donor tubing above the Y connector and disinfect site of phlebotomy.

4. If using blood pressure cuff, inflate to not more than 60 mm Hg.

5. Remove donor needle cover and accomplish phlebotomy.

6. Release clamp and ensure there is blood flow. Reduce pressure as required.

7. Slide the DCNG midway over the needle hub and securely tape DCNG to the donor's arm as close to the top of the DCNG as possible. Note: If blood flow is slow, slide DCNG away from the needle hub, adjust and re-engage DCNG. If repeated needle adjustment is necessary, slide DCNG away from the needle hub and re-engage at the end of blood collection.

8. Collect appropriate volume of blood into collection bag, as indicated on packaging.

Note: Mix blood and anticoagulant frequently during collection, for example, once every 45 seconds, and immediately after collection. If blood agitation device is used, follow manufacturer's operating instructions.

9. After required amount of blood has been collected, seal donor tubing between Y Sampling Site and the collection bag.*

10. For blood sampling, remove the Y Sampling Site needle cover. Ensure the protective sheath is in place over the sampling needle.

11. Fasten the vacuum tube holder on to the base of the sampling needle.

12. Collect blood samples into vacuum tubes.

13. Ensure the vacuum tubes are centered within the vacuum tube holder during sample collection.

14. Maintain forward pressure on the vacuum tubes during sample collection.

Note: After the last tube is collected, it is recommended that the vacuum tube holder be left in place.

15. After blood samples are collected, clamp donor tubing between the Y Sampling Site and DCNG, as close as possible to the DCNG.

16. Release any remaining pressure from the donor's arm.

17. DCNG must be held stationary while the needle is withdrawn into it. While holding sides of DCNG near the front, grasp the tubing below the clamp and pull the needle into the DCNG until it locks into place, and the needle hub engages the bottom of the DCNG.

18. Insert the DCNG into the vacuum tube holder. Note: It is recommended that the DCNG be inserted securely into the vacuum tube holder prior to discarding.

19. Seal donor tubing adjacent to DCNG.* Detach and discard needle, DCNG, Y Sampling Site and tubing.*

20. Strip tubing between seal and collection bag.

21. Continue to “Processing Instructions”, Section IV, Step 1.

II. BLOOD COLLECTION INSTRUCTION FOR SYSTEMS CONTAINING A SAMPLE DIVERSION POUCH WITH OR WITHOUT A PRE-ATTACHED SAMP LOK® VACUUM TUBE HOLDER

SPL UNCLASSIFIED SECTION

When using systems with a pre-attached Samp Lok vacuum tube holder, follow instructions as noted below, but refer to Section III when indicated to do so.

1. Load blood agitation device or suspend blood bag on donor scale and adjust donor scale to desired collection gross weight as per manufacturer's instructions.

2. Clamp donor tubing between the Donor Care® Needle Guard (DCNG) and Sampling Site.

3. Secure donor tubing above the Y connector and disinfect site of phlebotomy.

4. If using blood pressure cuff, inflate to not more than 60 mm Hg.

5. Remove donor needle cover and accomplish phlebotomy.

6. Release clamp and ensure there is blood flow.

7. Slide the DCNG midway over the needle hub and securely tape DCNG to the donor's arm as close to the top of the DCNG as possible. Note: If blood flow is slow, slide DCNG away from the needle hub, adjust and re-engage DCNG. If repeated needle adjustment is necessary, slide DCNG away from the needle hub and re-engage at the end of blood collection.

8. The donor blood will be automatically diverted to the sample diversion pouch. Once the sample diversion pouch is filled, close clamp immediately on tubing between the sample diversion pouch and Y connector. Warning: To avoid risk of air embolism to donor, do not squeeze sample diversion pouch while tubing is open.

9. Open snap-open closure between the Y connector and the collection bag to initiate blood collection. Reduce pressure as needed.

10. Permanently seal tubing between the sample diversion pouch and the Y connector to maintain sterility of the system prior to collecting blood samples.* Note: When using systems with a pre-attached Samp Lok vacuum tube holder, go to Section III.

11. For blood sampling, remove the Sampling Site needle cover. Ensure the protective sheath is in place over the sampling needle.

12. Fasten the vacuum tube holder on to the base of the sampling needle.

13. Position the sample diversion pouch downwards so that the air rises to the top of the pouch and away from the vacuum tube holder. Note: Drawing air into the vacuum tube may cause hemolysis.

14. Collect blood samples from the sample diversion pouch into vacuum tubes within approximately four minutes to avoid possible clot formation.

15. Ensure the vacuum tubes are centered within the vacuum tube holder during sample collection.

16. Maintain forward pressure on the vacuum tubes during sample collection.

Note: After the last tube is collected, it is recommended that the vacuum tube holder be left in place.

17. Collect appropriate volume of blood into collection bag as indicated on packaging.

Note: Mix blood and anticoagulant frequently during collection, for example, once every 45 seconds, and immediately after collection. If blood agitation device is used, follow manufacturer's operating instructions.

18. After required amount of blood has been collected, seal donor tubing close to snap open closure.* Note: If pre-filtration quality control is to be performed, leave an adequate length (~10 inches) of QC tubing containing anticoagulated blood attached to the collection bag.

19. Clamp donor tubing between the Y connector and DCNG, as close as possible to the DCNG.

20. Release any remaining pressure from the donor's arm.

21. DCNG must be held stationary while the needle is withdrawn into it. While holding sides of DCNG near the front, grasp the tubing below the clamp and pull the needle into the DCNG until it locks into place, and the needle hub engages the bottom of the DCNG.

22. Insert the DCNG into the vacuum tube holder, if desired. Note: It is recommended that the DCNG be inserted securely into the vacuum tube holder prior to discarding.

23. Seal donor tubing adjacent to DCNG.* Detach and discard needle, DCNG, sample diversion pouch and tubing.*

24. Strip tubing between seal and collection bag.

25. Continue to "Processing Instructions", Section IV, Step 1.

III. WHEN USING SYSTEMS WITH A PRE-ATTACHED SAMP LOK® VACUUM TUBE HOLDER

SPL UNCLASSIFIED SECTION

1. To collect blood samples, open lid from Samp Lok vacuum tube holder.

2. Open snap-open closure between sample diversion pouch and Samp Lok vacuum tube holder.

3. Position the sample diversion pouch downwards so that the air rises to the top of the pouch and away from the Samp Lok vacuum tube holder. Note: Drawing air into the vacuum tube may cause hemolysis.

4. Collect blood samples from the sample diversion pouch into vacuum tubes within approximately four minutes to avoid possible clot formation.

5. Ensure the vacuum tubes are centered within the Samp Lok vacuum tube holder during sample collection.

6. Maintain forward pressure on the vacuum tubes during sample collection.

7. The lid may be closed on the Samp Lok vacuum tube holder after sample collection.

8. Return to Section II, Step 17. Note: When collection of unit is complete and the donor needle is engaged in the DCNG, open the lid of the Samp Lok vacuum tube holder and insert the DCNG into the holder. Twist until it locks into place. An audible click will confirm that it is locked.

IV. Blood Filtration Instructions

SPL UNCLASSIFIED SECTION

  • Filter and process whole blood within 72 hours of collection.
  • Mix whole blood/anticoagulant gently and thoroughly.
  • 3. Place empty red cell storage bag on a horizontal surface.
  • 4. Ensure cap of blood recovery vent above the filter is tightly closed.
  • 5. Hang whole blood bag up to 60 inches (1.52 meters) above empty red cell storage bag and ensure filter is vertical. Note: The maximum head height should be 60 inches (1.52 meters) for whole blood filtered at room temperature and at 1—6 °C.
  • 6. Open snap-open closure of whole blood bag. Priming will occur automatically by gravity.
  • 7. After blood fills one side of the air vent below the filter, open snap-open closure immediately below the air vent to begin filtration.
  • 8. Allow blood to filter by gravity. Note: Do not apply mechanical or manual pressure to increase flow rate. Filtration is complete when whole blood bag is empty.
  • 9. Remove cap from blood recovery vent above the filter and allow the upstream side (non-printed side) of the filter to drain.
  • 10. When the upstream side of the filter is empty, clamp tubing below the filter. Note: Filtration times can be influenced by collection and processing conditions and biological variability of donors. Experimental data with some filter products indicate that a prolonged filtration can be an indication of sub-optimal leukocyte reduction.
  • 11. Seal tubing just below the snap-open closure.*
  • 12. Detach and discard collection bag and filter.* Note: Do not strip tubing prior to sealing tubing. If it is desired to strip blood from numbered tubing, do so only after tubing has been sealed close to the snap-open closure and detached.
  • 13. If desired, seal at or adjacent to "X" marks on the tubing to provide numbered segments of anticoagulated blood for typing or crossmatching.* Notes: If it is necessary to strip blood from numbered tubing for re-suspension, care should be taken when stripping is performed. Increased (mechanical) hemolysis has been associated with stripping when blood is cold and has a higher hematocrit. Do not strip forcefully or frequently against a snap-open closure.

V. PROCESSING INSTRUCTIONS

SPL UNCLASSIFIED SECTION

  • 1. Load whole blood and remaining bags into centrifuge cup, ensuring that the tubing stays in the top half of the cup.
  • 2. Centrifuge at conditions designed to produce desired components.
  • 3. Carefully remove the unit from the centrifuge and place the collection bag in the plasma expressor.
  • 4. Clamp tubing to extra satellite bag(s), if present.
  • 5. Gently apply expressor pressure.
  • 6. Open snap-open closure on collection bag and express plasma.
  • 7. After plasma is expressed, clamp tubing between satellite bag(s) whole blood bag and release expressor pressure.
  • 8. Seal tubing and detach the bag containing packed red cells, and set aside plasma for further processing.* Note: If pre-filtration quality control is to be performed, leave an adequate length (~10 inches) of QC tubing containing anticoagulated blood attached to the collection bag.
  • 9. Gently mix packed red cells.
  • 10. Store CPDA-1 preserved red blood cells at 1—6 °C for up to 35 days and use as indicated.

HOW SUPPLIED

HOW SUPPLIED SECTION

Each shipping case contains 6 foil envelopes. Within each foil envelope resides 3 units of the collection system. Each unit consists of a collection bag with 70 ml of CPDA-1 solution and two empty bags.

REFERENCES

REFERENCES SECTION

HAEMONETICS and THE Blood Management Company are trademarks of Haemonetics Corporation.

DonorCare and SampLok are registered trademarks of ITL Corporation, Canberra, Australia.

Manufactured for:

Haemonetics Corporation

400 Wood Road

Braintree, MA 02184, USA

By: Haemonetics Manufacturing Inc.

1630 Industrial Park Street

Covina, CA 91722, USA

www.haemonetics.com

800-537-2802

PACKAGE/LABEL PRINCIPAL DISPLAY PANEL

PACKAGE LABEL.PRINCIPAL DISPLAY PANEL

boxboxpouchpouch

DailyMed Pharmacologic Classes#

Class, Version, Type table
ClassVersionTypeEffective
ANHYDROUS CITRIC ACID Pharmacologic Class Indexing3Indexing - Pharmacologic Class20180813
SODIUM CATION Pharmacologic Class Indexing2Indexing - Pharmacologic Class20180813

DailyMed Product Concepts#

Product concept, Relation, Version table
Product conceptRelationVersionEffective
95250459-30e4-45c1-b08a-993044f49109Product name220250805
efd58dcf-540a-4531-8766-e713129ca6f2Product name120250307
4b7700e3-6e0e-45e5-9d22-6f754d61386eProduct name220250304
726062af-1135-4707-a1d7-57256991bbf9Product name220250226
a21a21e5-f36d-4c1f-8918-def921b8f7e3Product name320250124
fb7ab793-2c12-4079-b100-a64f73bef25aProduct name420240712
b0cbf770-6cc3-4aa4-9158-755110c2b9f7Product name220230717
cb89b86a-369e-4204-8910-ae797c503af3Product name120230706
dc7c5daa-021f-40dd-b00d-63982cb2067aProduct name120230426
6bd95106-a412-1dad-b9cc-4cb74bfb27ceProduct name220230315
f1e0dc7e-7d61-4625-b6d9-4894927679d7Product name220211025
9badc7be-250a-44ab-aa36-926af3f02679Product name120210527
08ffbcbf-26df-b99c-1dab-64fc4cfae89fProduct name520200925
e4870509-7e28-6f92-f0be-dd3dd889a9cfProduct name520200810
6e9d4a47-c464-b93e-aebc-ad7bffc9c4cdProduct name220190619
444b3e50-f226-46ef-bfca-2e7035d140cdProduct name120190611
816b97af-edc5-4060-aff1-b814bdbcad50Product name120190415
d4880ace-a516-8c47-8dfd-73a4ba5887d7Product name220180830
7cda52fc-125f-421c-8fea-bc1974370c49Product name220180703
419aab54-5d5a-4146-9453-026d4a9991beProduct name220170525
26acd337-b838-40ac-bcbc-05c3b81c8712Product name120170323
08ffbcbf-26df-b99c-1dab-64fc4cfae89fProduct name220160823
d5e51f11-ad28-caa4-4b49-4143974782adProduct name120150831
290f523a-f9db-9774-b5a9-e1f908ac1782Product name120150828
0ca1d589-929b-4b33-bc5b-1d84abdafa6aProduct name120150324
fc363c46-397b-4476-ac0f-70e43e8e4592Product name120150324
89dac932-b90a-4410-9ab1-84c53e57de25Product name120150316
19c71a3d-ed9c-166b-a7e5-38c250c35631Product name120140508
5668b646-cd56-c3c7-bdea-3f6b1a8840dbProduct name120140508
597b1cfd-7b31-5048-62be-ca5cd740da2fProduct name120140508
684ea482-2e8b-14af-5664-b98f1e402036Product name120140508
810ab97e-f109-f41c-7c83-6a652a9cbf43Product name120140508
87711080-88eb-65c5-b2dd-bf99e700a372Product name120140508
8dbefedf-0a0d-a224-5a3c-66dc9e11c2ddProduct name120140508
b060cc1a-fe47-856b-0b77-4448b5963de8Product name120140508
c6b65c52-69c7-df49-550a-a50c137f6218Product name120140508
ec2149b3-5c6d-5344-f757-c86411073075Product name120140508
ed912195-5da0-0f2f-6f4b-3ef17710cbe3Product name120140508

FDA-Initiated Inactive NDC Indexing#

NDC, Effective, Action table
NDCEffectiveActionDocumentIndexing SPLRelated label
53157-136-922025-01-30C16284748780-12cef2736-afbb-d83d-e063-dadaa90ab31fLeukotrap WB System, CPDA-1 Blood Collection Set with WBF2 Filter and Sampling System

DailyMed Package Descriptions#

Package NDC, Product, Description table
Package NDCProductDescriptionFormQuantityStrengthSPL version
53157-136-92Leukotrap WB System70 mL in 1 BAGSOLUTION704

DailyMed Dashboard NDC Coverage#

NDC, Dashboard title, SPL version table
NDCDashboard titleSPL versionValidationDashboard ZIP
53157-136LEUKOTRAP WB SYSTEM (CPDA-1 BLOOD BAG COLLECTION SYSTEM) SOLUTION [HAEMONETICS CORPORATION]4Legacy NDC, 1 package rows20200103_cd05ee40-f7de-4d1b-b701-648ff2dfdb3e.zip

DailyMed Socrata Ingredients#

Ingredient, Type, UNII table
IngredientTypeUNIISPL versionUploaded
ADENINEACTIVE INGREDIENTJAC85A21611
CITRIC ACID MONOHYDRATEACTIVE INGREDIENT2968PHW8QP1
DEXTROSE MONOHYDRATEACTIVE INGREDIENTLX22YL083G1
SODIUM CITRATEACTIVE INGREDIENT1Q73Q2JULR1
SODIUM PHOSPHATE, MONOBASIC, MONOHYDRATEACTIVE INGREDIENT593YOG76RN1
ADENINEACTIVE MOIETYJAC85A21611
ANHYDROUS CITRIC ACIDACTIVE MOIETYXF417D3PSL1
ANHYDROUS DEXTROSEACTIVE MOIETY5SL0G7R0OK1
SODIUM CATIONACTIVE MOIETYLYR4M0NH371

Products#

Every source-derived product name is available through these pages.

DailyMed product names page 1 of 1 · 11 matching rows.

NDC Codes#

Product NDC, Package NDC table
Product NDCPackage NDC
53157-13653157-136-92

Ingredients#

Every source-derived ingredient row is available through these pages.

DailyMed ingredient rows page 1 of 1 · 5 matching rows.

Source Document#

Source XML

openFDA label cross-check#

OpenFDA label data provides additional search and identifier links. DailyMed’s Structured Product Label is the canonical label on FDA.report. Matching records are deduplicated before they are shown below.

Matched openFDA labels page 1 of 1 · 1 matching rows.

Brand, Generic, Manufacturer table
BrandGenericManufacturerSPL set IDEffective dateAvailable safety fieldsJoin
9b2d6596-8ea5-7714-e053-2995a90a620fcd05ee40-f7de-4d1b-b701-648ff2dfdb3e2020-01-02Warnings, Adverse reactionsExact identifier
spl id: 9b2d6596-8ea5-7714-e053-2995a90a620f
spl set id: cd05ee40-f7de-4d1b-b701-648ff2dfdb3e